T1 - Ventricular assist device using a thoracotomy-based implant technique
T2 - Multi-Center Implantation of the HeartMate 3 in Subjects With Heart Failure Using Surgical Techniques Other Than Full Median Sternotomy (HM3 SWIFT)
AU - Gosev, Igor
AU - Pham, Duc Thinh
AU - Um, John Y.
AU - Anyanwu, Anelechi C.
AU - Itoh, Akinobu
AU - Kotkar, Kunal
AU - Takeda, Koji
AU - Naka, Yoshifumi
AU - Peltz, Matthias
AU - Silvestry, Scott C.
AU - Couper, Gregory
AU - Leacche, Marzia
AU - Rao, Vivek
AU - Sun, Benjamin
AU - Tedford, Ryan J.
AU - Mokadam, Nahush
AU - McNutt, Robert
AU - Crandall, Daniel
AU - Mehra, Mandeep R.
AU - Salerno, Christopher T.
N1 - Publisher Copyright:
© 2024 The Authors
PY - 2024
Y1 - 2024
N2 - Objectives: The HeartMate 3 (Abbott) left ventricular assist device provides substantial improvement in long-term morbidity and mortality in patients with advanced heart failure. The Implantation of the HeartMate 3 in Subjects With Heart Failure Using Surgical Techniques Other Than Full Median Sternotomy study compares thoracotomy-based implantation clinical outcomes with standard median sternotomy. Methods: We conducted a prospective, multicenter, single-arm study in patients eligible for HeartMate 3 implantation with thoracotomy-based surgical technique (bilateral thoracotomy or partial upper sternotomy with left thoracotomy). The composite primary end point was survival free of disabling stroke (modified Rankin score >3), or reoperation to remove or replace a malfunctioning device, or conversion to median sternotomy at 6-months postimplant (elective transplants were treated as a success). The primary end point (noninferiority, −15% margin) was assessed with >90% power compared with a propensity score-matched cohort (ratio 1:2) derived from the Multi-Center Study of MagLev Technology in Patients Undergoing Mechanical Circulatory Support Therapy With HeartMate 3 continued access protocol. Results: The study enrolled 102 patients between December 2020 and July 2022 in the thoracotomy-based arm at 23 North American centers. Follow-up concluded in December 2022. In the Implantation of the HeartMate 3 in Subjects With Heart Failure Using Surgical Techniques Other Than Full Median Sternotomy study group, noninferiority criteria was met (absolute between-group difference, −1.2%; Farrington Manning lower 1-sided 95% CI, −9.3%; P < .0025) and event-free survival was not different (85.0% vs 86.2%; hazard ratio, 1.01; 95% CI, 0.58-2.10). Length of stay with thoracotomy-based implant was longer (median, 20 vs 17 days; P = .03). No differences were observed for blood product utilization, adverse events (including right heart failure), functional status, and quality of life between cohorts. Conclusions: Thoracotomy-based implantation of the HeartMate 3 left ventricular assist device is noninferior to implantation via standard full sternotomy. This study supports thoracotomy-based implantation as an additional standard for surgical implantation of the HeartMate 3 left ventricular assist device.
AB - Objectives: The HeartMate 3 (Abbott) left ventricular assist device provides substantial improvement in long-term morbidity and mortality in patients with advanced heart failure. The Implantation of the HeartMate 3 in Subjects With Heart Failure Using Surgical Techniques Other Than Full Median Sternotomy study compares thoracotomy-based implantation clinical outcomes with standard median sternotomy. Methods: We conducted a prospective, multicenter, single-arm study in patients eligible for HeartMate 3 implantation with thoracotomy-based surgical technique (bilateral thoracotomy or partial upper sternotomy with left thoracotomy). The composite primary end point was survival free of disabling stroke (modified Rankin score >3), or reoperation to remove or replace a malfunctioning device, or conversion to median sternotomy at 6-months postimplant (elective transplants were treated as a success). The primary end point (noninferiority, −15% margin) was assessed with >90% power compared with a propensity score-matched cohort (ratio 1:2) derived from the Multi-Center Study of MagLev Technology in Patients Undergoing Mechanical Circulatory Support Therapy With HeartMate 3 continued access protocol. Results: The study enrolled 102 patients between December 2020 and July 2022 in the thoracotomy-based arm at 23 North American centers. Follow-up concluded in December 2022. In the Implantation of the HeartMate 3 in Subjects With Heart Failure Using Surgical Techniques Other Than Full Median Sternotomy study group, noninferiority criteria was met (absolute between-group difference, −1.2%; Farrington Manning lower 1-sided 95% CI, −9.3%; P < .0025) and event-free survival was not different (85.0% vs 86.2%; hazard ratio, 1.01; 95% CI, 0.58-2.10). Length of stay with thoracotomy-based implant was longer (median, 20 vs 17 days; P = .03). No differences were observed for blood product utilization, adverse events (including right heart failure), functional status, and quality of life between cohorts. Conclusions: Thoracotomy-based implantation of the HeartMate 3 left ventricular assist device is noninferior to implantation via standard full sternotomy. This study supports thoracotomy-based implantation as an additional standard for surgical implantation of the HeartMate 3 left ventricular assist device.
KW - HeartMate 3
KW - LVAD
KW - minimally invasive
KW - outcomes
KW - thoracotomy
UR - http://www.scopus.com/inward/record.url?scp=85189505035&partnerID=8YFLogxK
U2 - 10.1016/j.jtcvs.2024.02.013
DO - 10.1016/j.jtcvs.2024.02.013
M3 - Article
C2 - 38367698
AN - SCOPUS:85189505035
SN - 0022-5223
JO - Journal of Thoracic and Cardiovascular Surgery
JF - Journal of Thoracic and Cardiovascular Surgery
ER -