Abstract
Central-axis depth-dose data for 4-MV x rays, including tissue-maximum ratios, were measured for wedge fields. Comparison with corresponding open-field data revealed differences in magnitude which increased with depth, field size, and wedge thickness. However, phantom scatter correction factors for the wedge fields differed less than 1% from corresponding open-field factors. The differences in central-axis per cent depth doses between the two types of fields indicate beam hardening by the wedge filter. This study also implies that the derivation of tissue-maximum ratios from central-axis per cent depth is as valid for wedge as for open fields.
| Original language | English |
|---|---|
| Pages (from-to) | 789-792 |
| Number of pages | 4 |
| Journal | Radiology |
| Volume | 127 |
| Issue number | 3 |
| DOIs | |
| State | Published - 1978 |