T1 - Validation of creatinine assays utilizing HPLC and IDMS traceable standards in sera of children
AU - Schwartz, George J.
AU - Kwong, Tai
AU - Erway, Brian
AU - Warady, Bradley
AU - Sokoll, Lori
AU - Hellerstein, Stanley
AU - Dharnidharka, Vikas
AU - Furth, Susan
AU - Muñoz, Alvaro
PY - 2009
Y1 - 2009
N2 - The purpose of this study was to validate serum creatinine (SCr) concentrations assayed in the Central Biochemistry Laboratory of the National Institutes of Health (NIH)-funded Chronic Kidney Disease in Children (CKiD) study utilizing an enzymatic assay (Siemens Advia 2400) against a method traceable to reference isotope dilution mass spectroscopy (IDMS) developed by the National Institute of Standards and Technology (NIST). High-performance liquid chromatography (HPLC) measured SCr after external validation utilizing IDMS-based standard reference materials. Sera from the first 201 subjects enrolled in CKiD were analyzed and compared for creatinine concentration by enzymatic and HPLC methods. Fifty "normal" pediatric sera were subsequently analyzed. Finally, a "pediatric" reference standard was prepared and examined for accuracy and precision. Enzymatic SCr concentrations (median 1.4 mg/dl) of CKiD subjects were well correlated with HPLC (r=0.984) but were slightly higher (+7%; p<0.001). Agreement was poorer at lower SCr (median 0.4 mg/dl) when using samples from normal children and the "pediatric" reference standard. However, the Roche enzymatic assay was comparable with HPLC in accuracy and precision. Referring physicians should be aware of the accuracy and reproducibility of their laboratory's SCr assay. Our enzymatic assay agreed well with HPLC in CKiD subjects with elevated SCr. We suggest that NIST develop a pediatric SCr standard reference material for use by assay manufacturers to improve accuracy and precision of assays at the low SCr levels observed in most pediatric patients.
AB - The purpose of this study was to validate serum creatinine (SCr) concentrations assayed in the Central Biochemistry Laboratory of the National Institutes of Health (NIH)-funded Chronic Kidney Disease in Children (CKiD) study utilizing an enzymatic assay (Siemens Advia 2400) against a method traceable to reference isotope dilution mass spectroscopy (IDMS) developed by the National Institute of Standards and Technology (NIST). High-performance liquid chromatography (HPLC) measured SCr after external validation utilizing IDMS-based standard reference materials. Sera from the first 201 subjects enrolled in CKiD were analyzed and compared for creatinine concentration by enzymatic and HPLC methods. Fifty "normal" pediatric sera were subsequently analyzed. Finally, a "pediatric" reference standard was prepared and examined for accuracy and precision. Enzymatic SCr concentrations (median 1.4 mg/dl) of CKiD subjects were well correlated with HPLC (r=0.984) but were slightly higher (+7%; p<0.001). Agreement was poorer at lower SCr (median 0.4 mg/dl) when using samples from normal children and the "pediatric" reference standard. However, the Roche enzymatic assay was comparable with HPLC in accuracy and precision. Referring physicians should be aware of the accuracy and reproducibility of their laboratory's SCr assay. Our enzymatic assay agreed well with HPLC in CKiD subjects with elevated SCr. We suggest that NIST develop a pediatric SCr standard reference material for use by assay manufacturers to improve accuracy and precision of assays at the low SCr levels observed in most pediatric patients.
KW - Enzymatic creatinine assay
KW - Glomerular filtration rate
KW - High-performance liquid chromatography
KW - Isotope dilution mass spectroscopy
KW - Jaffe creatinine assay
KW - Pediatric standards
UR - http://www.scopus.com/inward/record.url?scp=59849089612&partnerID=8YFLogxK
U2 - 10.1007/s00467-008-0957-0
DO - 10.1007/s00467-008-0957-0
M3 - Article
C2 - 18769945
AN - SCOPUS:59849089612
SN - 0931-041X
VL - 24
SP - 113
EP - 119
JO - Pediatric Nephrology
JF - Pediatric Nephrology
IS - 1
ER -