TY - JOUR
T1 - Use of pH 9.5 Tris-HCl buffer containing 5% urea for antigen retrieval immunohistochemistry
AU - Shan-Rong-Shi,
AU - Cote, R. J.
AU - Young, L.
AU - Imam, S. A.
AU - Taylor, C. R.
PY - 1996
Y1 - 1996
N2 - Successful antigen retrieval (AR) immunohistochemistry is dependent on the temperature, heating time, and pH value of the AR solutions. There is no single standardized AR solution, however, that is suitable for all antibodies 'routinely' used in surgical pathology for immunostaining archival tissue sections. We tested a variety of AR solutions varying in pH value, chemical composition, and molarity. Based upon preliminary results, we compared three AR solutions: 0.1 M Tris-HCl buffer, pH 9.5, containing 5% urea, 0.1 M Tris- HCl buffer pH 9.5 without urea, and citrate buffer, pH 6.0. Each AR solution was tested with a panel of 34 antibodies using microwave heating for antigen retrieval. The heating conditions were standardized at 10 min and an automated stainer was used to standardize the immunostaining method. The Tris-HC1 containing urea was superior to pH 6.0 citrate buffer for 22 antibodies. In 12 cases, Tris-HCl with urea was also superior to Tris-HCl alone. In 12 cases, the intensity was similar for all three retrieval solutions. The staining obtained with Tris-HCl with urea was equal to or better than with pH 6.0 citrate buffer in all cases. The Tris-HC1 with urea solution is satisfactory for AR of most antibodies employed in routine surgical pathology.
AB - Successful antigen retrieval (AR) immunohistochemistry is dependent on the temperature, heating time, and pH value of the AR solutions. There is no single standardized AR solution, however, that is suitable for all antibodies 'routinely' used in surgical pathology for immunostaining archival tissue sections. We tested a variety of AR solutions varying in pH value, chemical composition, and molarity. Based upon preliminary results, we compared three AR solutions: 0.1 M Tris-HCl buffer, pH 9.5, containing 5% urea, 0.1 M Tris- HCl buffer pH 9.5 without urea, and citrate buffer, pH 6.0. Each AR solution was tested with a panel of 34 antibodies using microwave heating for antigen retrieval. The heating conditions were standardized at 10 min and an automated stainer was used to standardize the immunostaining method. The Tris-HC1 containing urea was superior to pH 6.0 citrate buffer for 22 antibodies. In 12 cases, Tris-HCl with urea was also superior to Tris-HCl alone. In 12 cases, the intensity was similar for all three retrieval solutions. The staining obtained with Tris-HCl with urea was equal to or better than with pH 6.0 citrate buffer in all cases. The Tris-HC1 with urea solution is satisfactory for AR of most antibodies employed in routine surgical pathology.
KW - antibody
KW - antigen
KW - antigen retrieval
KW - immunohistochemistry
KW - microwave
KW - paraffin section
UR - http://www.scopus.com/inward/record.url?scp=0029740847&partnerID=8YFLogxK
U2 - 10.3109/10520299609117158
DO - 10.3109/10520299609117158
M3 - Article
C2 - 8874856
AN - SCOPUS:0029740847
SN - 1052-0295
VL - 71
SP - 190
EP - 196
JO - Biotechnic and Histochemistry
JF - Biotechnic and Histochemistry
IS - 4
ER -