T1 - Trimodal wireless intramuscular device detects muscle pressure, flow, and oxygenation changes in porcine model of lower extremity compartment syndrome
AU - Westman, Amanda M.
AU - Ribaudo, Joseph
AU - Seo, Seung Gi
AU - Moritz, William
AU - Tatman, Lauren M.
AU - Jin, Sung Hun
AU - Kim, Seungyeob
AU - Oh, Seyong
AU - Rogers, John A.
AU - Pet, Mitchell A.
N1 - Publisher Copyright:
© The Author(s), under exclusive licence to Springer-Verlag France SAS, part of Springer Nature 2024.
PY - 2024
Y1 - 2024
N2 - Purpose: Compartment syndrome remains difficult to diagnose early in its clinical course. Pressure transducer catheters have been used to directly measure intracompartmental pressure (ICP), but this method is unreliable, with a false positive rate of 35%. We have previously used intramuscular near infrared spectroscopy to detect changes in tissue oxygen saturation (StO2) in response to increasing ICP using a novel implantable probe. However, measuring StO2 may not be sufficient to identify CS in the clinical setting. The pathophysiology of CS consists of increased ICP, leading to decreased tissue perfusion, and resulting in reduced tissue oxygenation. More clinically useful information may come from the integration of multiple data streams to aid in the diagnosis of CS. In this study, we present a novel, intramuscular probe capable of simultaneous measurement of ICP, StO2, and microvascular blood flow in a porcine model of ACS. Methods: Proof of concept for this device is demonstrated in a porcine lower extremity balloon compression model of ACS. Pressure was maintained for 20 min (short-term) or 3 h (long-term) before the balloon volume was removed. Results: In both short- and long-term experiments, as ICP increased with increasing balloon volume, the novel multimodal sensor simultaneously and reliably detected pressure elevation and corresponding reversible reductions in microvascular flow rate and tissue oxygenation. Conclusion: This novel trimodal device simultaneously measured the elevated ICP, decreased perfusion, and tissue ischemia of evolving ACS, substantiating our basic understanding of CS pathophysiology.
AB - Purpose: Compartment syndrome remains difficult to diagnose early in its clinical course. Pressure transducer catheters have been used to directly measure intracompartmental pressure (ICP), but this method is unreliable, with a false positive rate of 35%. We have previously used intramuscular near infrared spectroscopy to detect changes in tissue oxygen saturation (StO2) in response to increasing ICP using a novel implantable probe. However, measuring StO2 may not be sufficient to identify CS in the clinical setting. The pathophysiology of CS consists of increased ICP, leading to decreased tissue perfusion, and resulting in reduced tissue oxygenation. More clinically useful information may come from the integration of multiple data streams to aid in the diagnosis of CS. In this study, we present a novel, intramuscular probe capable of simultaneous measurement of ICP, StO2, and microvascular blood flow in a porcine model of ACS. Methods: Proof of concept for this device is demonstrated in a porcine lower extremity balloon compression model of ACS. Pressure was maintained for 20 min (short-term) or 3 h (long-term) before the balloon volume was removed. Results: In both short- and long-term experiments, as ICP increased with increasing balloon volume, the novel multimodal sensor simultaneously and reliably detected pressure elevation and corresponding reversible reductions in microvascular flow rate and tissue oxygenation. Conclusion: This novel trimodal device simultaneously measured the elevated ICP, decreased perfusion, and tissue ischemia of evolving ACS, substantiating our basic understanding of CS pathophysiology.
KW - Compartment syndrome
KW - Intramuscular
KW - Oxygenation monitoring
KW - Perfusion monitoring
KW - Pressure monitoring
UR - http://www.scopus.com/inward/record.url?scp=85195308755&partnerID=8YFLogxK
U2 - 10.1007/s00590-024-04023-4
DO - 10.1007/s00590-024-04023-4
M3 - Article
C2 - 38847913
AN - SCOPUS:85195308755
SN - 1633-8065
JO - European Journal of Orthopaedic Surgery and Traumatology
JF - European Journal of Orthopaedic Surgery and Traumatology
ER -