Abstract
We have developed an endoscopic OCT system based on an FDML laser. In vivo threedimensional imaging at 100 kHz with 9 × 7 μm (transverse by axial) resolution is demonstrated in the rabbit gastrointestinal tract.
Original language | English |
---|---|
Pages | BWF8 |
DOIs | |
State | Published - 2008 |
Event | Biomedical Optics, BIOMED 2008 - St. Petersburg, FL, United States Duration: Mar 16 2008 → Mar 19 2008 |
Conference
Conference | Biomedical Optics, BIOMED 2008 |
---|---|
Country/Territory | United States |
City | St. Petersburg, FL |
Period | 03/16/08 → 03/19/08 |