Technical Note: Dosimetric characterization of the dynamic beam flattening MLC sequence on a ring shaped, Jawless Linear Accelerator with double stacked MLC

Douglas Bollinger, Eric Laugeman, Taoran Li, Jessica Hilliard, Ana Heermann, Hyun Kim, Geoff Hugo, Sasa Mutic, Lei Dong, Bin Cai

Research output: Contribution to journalArticlepeer-review

6 Scopus citations

Abstract

Purpose: To characterize the dosimetric features and limitations of the dynamic beam flattening (DBF) on the Halcyon 2.0 linear accelerator (Varian Medical Systems). Methods: A predefined multi-leaf collimator (MLC) sequence was introduced and used to flatten the 6 MV flattening filter free (FFF) beam on the Halcyon 2.0. Dosimetric characterizations of the flattened beams, including beam flatness, symmetry, percent depth dose (PDD), output factor and MU linearity, were investigated. Flatness and symmetry were obtained from profile measurements with both radiographic films (EDR2) and a two dimensional ion-chamber array (IC Profiler, Sun Nuclear Corporation). MU linearity, output factors, and PDDs were measured in a water tank with a CC13 ion chamber (Scanditronix Wellhöfer, Nuremburg, Germany). In addition, the effect of the DBF sequence on 3D plan quality was evaluated by creating DBF plans for a 4-field box rectum and an AP/PA spine plan. Patient specific QA was performed on these plans. Results: At 100 cm SSD and 10 cm depth, a flatness of <3% was observed on both transversal and radial profiles for all square field sizes ≥10 cm with DBF. For both larger and smaller field sizes the flatness showed a tendency to increase as the fields got bigger or smaller, respectively. Similar trends in flatness were observed at all depths measured. All measured output factors for square field sizes ≥5 cm were within 1% of the TPS prediction. Linearity was ≤2.02% for all measurements. For both treatment sites, the MD judged the plans created for the Halcyon without the use of DBF not to be clinically acceptable, however considered both the TrueBeam plan and the Halcyon plan with the DBF sequence to be clinically acceptable. Conclusions: The DBF sequence on the Halcyon and its characteristics were investigated. The analysis indicates that the DBF sequence can be used on the Halcyon to generate clinically acceptable 3D treatment plans.

Original languageEnglish
Pages (from-to)948-957
Number of pages10
JournalMedical physics
Volume47
Issue number3
DOIs
StatePublished - Mar 1 2020

Keywords

  • dynamic beam flattening
  • halcyon
  • photon radiotherapy

Fingerprint

Dive into the research topics of 'Technical Note: Dosimetric characterization of the dynamic beam flattening MLC sequence on a ring shaped, Jawless Linear Accelerator with double stacked MLC'. Together they form a unique fingerprint.

Cite this