Surface and buildup dose characteristics for 6, 10, and 18 MV photons from an Elekta Precise linear accelerator.

Eric E. Klein, Jacqueline Esthappan, Zuofeng Li

Research output: Contribution to journalArticlepeer-review

36 Scopus citations

Abstract

Understanding head scatter characteristics of photon beams is vital to properly commission treatment planning (TP) algorithms. Simultaneously, having definitive surface and buildup region dosimetry is important to optimize bolus. The Elekta Precise linacs have unique beam flattening filter configurations for each photon beam (6, 10, and 18 MV) in terms of material and location. We performed a comprehensive set of surface and buildup dose measurements with a thin window parallel-plate (PP) chamber to examine effects of field size (FS), source-to-skin distance (SSD), and attenuating media. Relative ionization data were converted to fractional depth dose (FDD) after correcting for bias effects and using the Gerbi method to account for chamber characteristics. Data were compared with a similar vintage Varian linac. At short SSDs the surface and buildup dose characteristics were similar to published data for Varian and Elekta accelerators. The FDD at surface (FDD(0)) for 6, 10, and 18 MV photons was 0.171, 0.159, and 0.199, respectively, for a 15x15 cm2, 100 cm SSD field. A blocking tray increased FDD(0) to 0.200, 0.200, and 0.256, while the universal wedge decreased FDD(0) to 0.107, 0.124, and 0.176. FDD(0) increased linearly with FS (approximately 1.16%/cm). FDD(0) decreased exponentially for 10 and 18 MV with increasing SSD. However, the 6 MV FDD(0) actually increased slightly with increasing SSD. This is likely due to the unique distal flattening filter for 6 MV. The measured buildup curves have been used to optimize TP calculations and guide bolus decisions. Overall the FDD(0) and buildup doses were very similar to published data. Of interest were the relatively low 10 MV surface doses, and the 6 MV FDD(0)'s dependence on SSD.

Original languageEnglish
Pages (from-to)1-7
Number of pages7
JournalJournal of applied clinical medical physics / American College of Medical Physics
Volume4
Issue number1
DOIs
StatePublished - 2003

Fingerprint

Dive into the research topics of 'Surface and buildup dose characteristics for 6, 10, and 18 MV photons from an Elekta Precise linear accelerator.'. Together they form a unique fingerprint.

Cite this