SU‐FF‐J‐79: Quantification of Pitch Angle Deviations in Patients Receiving IMRT for Prostate Cancer, Without the Use of Fiducial Seed Markers

J. Perks, J. Lehmann, A. Rioux, S. Semon, J. Purdy, S. Vijayakumar

Research output: Contribution to journalArticlepeer-review

Abstract

Purpose: To demonstrate the strength of kV cone beam CT (CBCT) over MV portal imaging in the detection and quantification of pitch errors in prostate patients without use of internal fiducial markers. Additionally the cause and subsequent correction of large pitch variations was investigated as the daily use of CBCT had highlighted a small number of patients with deviations is set‐up of more than three degrees. Method and Materials: By using CBCT six variables can be corrected — three translations in the orthogonal axes as well as pitch, yaw, and roll. Five patients and RANDO phantom were studied, with significant differences being seen in pitch angle for two patients in the same direction as couch sag. Slight alterations were made in the position of the patients' knees and ankles in an attempt to reduce the pitch deviations. For all patients right lateral MV portal images and corresponding digitally reconstructed radiographs (DRR) were shown to the attending physician and a senior therapist with the question to determine if pitch greater than 3 degrees could be determined. Results: The deviation needed to create a pitch angle of 5 degrees in the RANDO phantom is significant so couch sag does not explain the deviations seen in a small population of the prostate patients imaged with CBCT. MV images were judged as showing insufficient information to determine pitch and Conclusion: Our clinical practice is that when any angular deviation over 3 degrees is observed in a patient set‐up the attending physician is called to decide how to proceed — treat or set‐up again. The phantom work demonstrated that a pitch variation in a patient set‐up cannot be attributed to couch sag and patient studies show that a few centimeters difference in knee separation lead to up to 3 degrees increase in pitch angle.

Original languageEnglish
Number of pages1
JournalMedical physics
Volume34
Issue number6
DOIs
StatePublished - Jun 2007

Fingerprint

Dive into the research topics of 'SU‐FF‐J‐79: Quantification of Pitch Angle Deviations in Patients Receiving IMRT for Prostate Cancer, Without the Use of Fiducial Seed Markers'. Together they form a unique fingerprint.

Cite this