SIMS Data Correction Procedure for Quasi-Simultaneous Arrival (QSA) Under-counting and Ramifications of Misapplication

Clive Jones, David A. Fike

Research output: Contribution to journalArticlepeer-review

1 Scopus citations

Abstract

Isotope geochemistry requires isotope ratios measured using secondary ion mass spectrometry (SIMS) to be made with optimal precision and accuracy. Under some analytical conditions when using electron multiplier detectors, secondary ions may be under-counted because of quasi-simultaneous arrival (QSA) at the first dynode. The relative magnitude of the associated QSA correction to raw measured isotopic ratios can be up to seventy permil or more. Therefore, not applying the correction, or misapplication of it could lead to significant inaccuracies in published isotope ratio data. Examples and ramifications of the latter are described in addition to a straightforward procedure for QSA under-counting correction.

Original languageEnglish
Pages (from-to)613-618
Number of pages6
JournalGeostandards and Geoanalytical Research
Volume48
Issue number3
DOIs
StatePublished - Sep 2024

Keywords

  • QSA correction method
  • SIMS
  • quasi-simultaneous arrival
  • secondary ion mass spectrometry