Abstract
We developed second-generation (G2) optical-resolution photoacoustic microscopy (OR-PAM). Incorporation of a novel acoustic detection scheme improved upon the sensitivity of our first-generation (G1) system by 18.4 dB, deepening the in vivo tissue penetration to 1.2mm at 570nm. Moreover, translating the imaging head instead of the living object accelerated the scanning speed by a factor of 5, widening the field of view within the same acquisition time. Mouse ears, as well as mouse brains with intact craniums, were imaged in vivo in both total concentration and oxygen saturation of hemoglobin.
| Original language | English |
|---|---|
| Pages (from-to) | 1134-1136 |
| Number of pages | 3 |
| Journal | Optics Letters |
| Volume | 36 |
| Issue number | 7 |
| DOIs | |
| State | Published - Apr 1 2011 |
Fingerprint
Dive into the research topics of 'Second-generation optical-resolution photoacoustic microscopy with improved sensitivity and speed'. Together they form a unique fingerprint.Cite this
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver