Retrieval and analysis of intramedullary rods: A follow-up study

  • S. D. Cook
  • , R. L. Barrack
  • , E. Renz
  • , R. C. Anderson
  • , A. F. Harding
  • , R. J. Haddad
  • , M. Milicic

Research output: Contribution to journalArticlepeer-review

6 Scopus citations

Abstract

From 1980 to 1983, 16 AO intramedullary femoral rods were retrieved and analyzed clinically, radiographically, and metallurgically. Cracking and/or fracture was observed in four of the 16 specimens. All cracks occurred at the point of maximum stress at the end of the slot. In two cases a structurally weaker weld zone coincided with this location. The weld zone and slot were also found to coincide in five of the uncracked rods. Two of the four cracked rods were removed because of pain, while only three of the 12 uncracked rods caused pain. Structural and material characteristics (i.e., the location of the weld at a point of maximum stress and significant variability in microstructure), rather than surgical technique or time in situ, were found to be responsible for the implant mechanical failures. Improvements should be made in fabrication techniques and material properties. For the present, careful assessment of a painful intramedullary rod and routine removal after healing are advisable.

Original languageEnglish
Pages (from-to)269-273
Number of pages5
JournalClinical orthopaedics and related research
VolumeNO. 191
StatePublished - Dec 1 1984

Fingerprint

Dive into the research topics of 'Retrieval and analysis of intramedullary rods: A follow-up study'. Together they form a unique fingerprint.

Cite this