Introduction Screw loosening, a common complication of patients undergoing a semi-rigid fixation, is associated with the radiological finding of lucency within bone surrounding the implanted screw. Case report A clinical and radiologic review is presented of a patient who underwent rigid fixation with L4-5 interbody fusion and L3-4 posterolateral dynamic stabilization with Zimmer Dynamic-to-Optima (DTO) hybrid fusion to non-fusion system (Zimmer Spine, Minneapolis, MN) with hydroxyapatite coated pedicle screws that exhibited lucency on radiographic studies which subsequently disappeared. An L4-5 interbody fusion and L3-4 posterolateral dynamic stabilization was performed. The patient did well clinically postoperatively. At 4 months postoperatively, X-rays of the lumbar spine demonstrated the appearance of lucency around the left L3 pedicle screw, which began to resolve at 10 months, and continued to resolve on subsequent CT scan 1 year postoperatively. The patient's clinical status continued to improve, with no complications. Conclusions The development of radiolucent lines around pedicle screws several months postoperatively may be due to loosening of the screws or due to a stress response at the bone-screw interface. The subsequent disappearance of this finding would suggest the ongrowth of bone at the osteoconductive hydroxyapatite coating or the protection of the bone-screw interface afforded by dynamic stabilization.

Original languageEnglish
Pages (from-to)71-75
Number of pages5
JournalInterdisciplinary Neurosurgery: Advanced Techniques and Case Management
Volume9
DOIs
StatePublished - Sep 1 2017

Keywords

  • Dynamic stabilization
  • Hydroxyapatite
  • Lumbar
  • Radiolucency
  • Spinal surgery

Fingerprint

Dive into the research topics of 'Resolution of screw radiolucency following instrumentation with a hybrid rigid to dynamic stabilization system'. Together they form a unique fingerprint.

Cite this