TY - JOUR
T1 - Real-World Complications of the SpaceOAR Hydrogel Spacer
T2 - A Review of the Manufacturer and User Facility Device Experience Database
AU - Fernandez, Adrian M.
AU - Jones, Charles P.
AU - Patel, Hiren V.
AU - Ghaffar, Umar
AU - Hakam, Nizar
AU - Li, Kevin D.
AU - Nabavizadeh, Behnam
AU - Breyer, Benjamin N.
N1 - Publisher Copyright:
© 2023 Elsevier Inc.
PY - 2024/1
Y1 - 2024/1
N2 - Objective: To characterize adverse events related to use of the perirectal spacing agent SpaceOAR, we examined the Manufacturer and User Facility Device Experience (MAUDE) database. Methods: The MAUDE database was queried for “SpaceOAR” and “Augmenix” from June 2015 (when SpaceOAR was approved by the Food and Drug Administration) to October 2022. Reports were reviewed for adverse events (AEs), operative procedures performed because of the AE, and changes to the radiation plan. AEs were categorized using Common Terminology Criteria for Adverse Events (CTCAE), version 5.0. Results: Six hundred fifty-four reports were reviewed. Eighty-four were excluded and 4 reports reviewed 2 separate cases of SpaceOAR administration. Five hundred seventy-four cases were ultimately included. Three deaths were reported (0.5% of all AEs). One point six percent of cases represented CTCAE grade 4 injuries (life-threatening consequences; urgent intervention indicated), 15.9% grade 3 (severe but not immediately life-threatening; hospitalization), 24.2% grade 2 (moderate; local/noninvasive intervention), and 57% of events were CTCAE grade 1 (mild; asymptomatic or mild symptoms). Bowel diversion occurred in 29 cases (9%). Conclusion: Both asymptomatic (n = 311) and debilitating (n = 12) complications of SpaceOAR hydrogel use were identified. Death, gel embolization, anaphylaxis, rectal ulcerations, and infections requiring bowel or urinary diversions were among the complications reviewed. Providers should consider these potential complications before perirectal spacer administration and during patient counseling.
AB - Objective: To characterize adverse events related to use of the perirectal spacing agent SpaceOAR, we examined the Manufacturer and User Facility Device Experience (MAUDE) database. Methods: The MAUDE database was queried for “SpaceOAR” and “Augmenix” from June 2015 (when SpaceOAR was approved by the Food and Drug Administration) to October 2022. Reports were reviewed for adverse events (AEs), operative procedures performed because of the AE, and changes to the radiation plan. AEs were categorized using Common Terminology Criteria for Adverse Events (CTCAE), version 5.0. Results: Six hundred fifty-four reports were reviewed. Eighty-four were excluded and 4 reports reviewed 2 separate cases of SpaceOAR administration. Five hundred seventy-four cases were ultimately included. Three deaths were reported (0.5% of all AEs). One point six percent of cases represented CTCAE grade 4 injuries (life-threatening consequences; urgent intervention indicated), 15.9% grade 3 (severe but not immediately life-threatening; hospitalization), 24.2% grade 2 (moderate; local/noninvasive intervention), and 57% of events were CTCAE grade 1 (mild; asymptomatic or mild symptoms). Bowel diversion occurred in 29 cases (9%). Conclusion: Both asymptomatic (n = 311) and debilitating (n = 12) complications of SpaceOAR hydrogel use were identified. Death, gel embolization, anaphylaxis, rectal ulcerations, and infections requiring bowel or urinary diversions were among the complications reviewed. Providers should consider these potential complications before perirectal spacer administration and during patient counseling.
UR - http://www.scopus.com/inward/record.url?scp=85174742159&partnerID=8YFLogxK
U2 - 10.1016/j.urology.2023.09.016
DO - 10.1016/j.urology.2023.09.016
M3 - Article
C2 - 37774851
AN - SCOPUS:85174742159
SN - 0090-4295
VL - 183
SP - 157
EP - 162
JO - Urology
JF - Urology
ER -