TY - JOUR
T1 - Radiofrequency ablation with a new perfused-cooled electrode using a single pump
T2 - An experimental study in ex vivo bovine liver
AU - Seung, Kwon Kim
AU - Seo, Jung Wook
N1 - Copyright:
Copyright 2008 Elsevier B.V., All rights reserved.
PY - 2005/12
Y1 - 2005/12
N2 - The purpose of this study was to assess the efficacy of a new perfused-cooled electrode that uses a single pump for creating a large ablation zone in explanted bovine liver. This was done by comparing with the radiofrequency (RF) ablation zones that were created with a monopolar cooled electrode to the RF ablation zones that were created by the new perfused-cooled electrode. We developed a new perfused-cooled electrode that uses a single pump by modifying a 17-gauge cooled electrode (Radionics) with a 2.5-cm outer metallic sheath (15-gauge) in order to allow use of the internal cooling water (5.85 % hypertonic saline) for the infused saline. Thirty ablation zones were created in explanted bovine livers (12-min ablation cycle; pulsed technique; 2000 mA, maximum) with three different regimens: group A, RF ablation with the 17-gauge cooled electrode; group B, RF ablation with the 15-gauge cooled electrode; group C, RF ablation with the perfused-cooled electrode. T2-weighted magnetic resonance (MR) imaging was obtained immediately after RF ablation for calculating volumes of the ablation zone. Following MR imaging, the ablation zones were excised and measured for transverse diameters and vertical diameters. The transverse diameter, vertical diameter, and the calculated volumes of the ablation zones on MR imaging were compared among the groups. Ablation zones created with the perfused-cooled electrode (group C) were significantly larger than those created with the 17-gauge cooled electrode (group A) and the 15-gauge cooled electrode (group B) according to the transverse diameter and vertical diameter on the gross specimens (p < 0.05): 3.6 ± 0.38 cm and 4.4 ± 0.20 cm in group A, 3.7 ± 0.08 cm and 4.6 ± 0.16 cm in group B, and 5.4 ± 0.65 cm and 6.0 ± 0.56 cm in group C, respectively. On the MR imaging, the calculated volumes of the ablation zones in group C were significantly larger than those in groups A and B (p < 0.05): 23.1 ± 8.7 cm3 in group A, 28.9 ± 5.7 cm3 in group B, and 80.0 ± 34 cm3 in group C, respectively. A new perfused-cooled electrode using a single pump could efficiently increase the size of the ablation zone in liver compared with a monopolar cooled electrode, and this was due to its simultaneous use of internal cooling and saline infusion.
AB - The purpose of this study was to assess the efficacy of a new perfused-cooled electrode that uses a single pump for creating a large ablation zone in explanted bovine liver. This was done by comparing with the radiofrequency (RF) ablation zones that were created with a monopolar cooled electrode to the RF ablation zones that were created by the new perfused-cooled electrode. We developed a new perfused-cooled electrode that uses a single pump by modifying a 17-gauge cooled electrode (Radionics) with a 2.5-cm outer metallic sheath (15-gauge) in order to allow use of the internal cooling water (5.85 % hypertonic saline) for the infused saline. Thirty ablation zones were created in explanted bovine livers (12-min ablation cycle; pulsed technique; 2000 mA, maximum) with three different regimens: group A, RF ablation with the 17-gauge cooled electrode; group B, RF ablation with the 15-gauge cooled electrode; group C, RF ablation with the perfused-cooled electrode. T2-weighted magnetic resonance (MR) imaging was obtained immediately after RF ablation for calculating volumes of the ablation zone. Following MR imaging, the ablation zones were excised and measured for transverse diameters and vertical diameters. The transverse diameter, vertical diameter, and the calculated volumes of the ablation zones on MR imaging were compared among the groups. Ablation zones created with the perfused-cooled electrode (group C) were significantly larger than those created with the 17-gauge cooled electrode (group A) and the 15-gauge cooled electrode (group B) according to the transverse diameter and vertical diameter on the gross specimens (p < 0.05): 3.6 ± 0.38 cm and 4.4 ± 0.20 cm in group A, 3.7 ± 0.08 cm and 4.6 ± 0.16 cm in group B, and 5.4 ± 0.65 cm and 6.0 ± 0.56 cm in group C, respectively. On the MR imaging, the calculated volumes of the ablation zones in group C were significantly larger than those in groups A and B (p < 0.05): 23.1 ± 8.7 cm3 in group A, 28.9 ± 5.7 cm3 in group B, and 80.0 ± 34 cm3 in group C, respectively. A new perfused-cooled electrode using a single pump could efficiently increase the size of the ablation zone in liver compared with a monopolar cooled electrode, and this was due to its simultaneous use of internal cooling and saline infusion.
KW - Alcohol ablation- Experimental study
KW - Liver interventional procedures
KW - Liver, interventional procedure
KW - Radiofrequency ablation
UR - http://www.scopus.com/inward/record.url?scp=27744539216&partnerID=8YFLogxK
U2 - 10.1007/s00270-005-0047-7
DO - 10.1007/s00270-005-0047-7
M3 - Article
C2 - 16187149
AN - SCOPUS:27744539216
SN - 0174-1551
VL - 28
SP - 779
EP - 788
JO - CardioVascular and Interventional Radiology
JF - CardioVascular and Interventional Radiology
IS - 6