TY - JOUR
T1 - Racial discrepancy in pulse oximeter accuracy in preterm infants
AU - Vesoulis, Zachary
AU - Tims, Anna
AU - Lodhi, Hafsa
AU - Lalos, Natasha
AU - Whitehead, Halana
N1 - Publisher Copyright:
© 2021, The Author(s), under exclusive licence to Springer Nature America, Inc.
PY - 2022/1
Y1 - 2022/1
N2 - Objective: Pulse oximetry is commonly used in Neonatology, however recent adult data suggest racial disparity in accuracy, with overestimation of oxygen saturation for Black patients. Study design: Black and White infants <32 weeks gestation underwent simultaneous arterial blood gas and pulse oximetry measurement. Error by race was examined using mean bias, Arms, Bland–Altman, and linear/non-linear analysis. Results: A total of 294 infants (124 Black, 170 White) were identified with mean GA of 25.8 ± 2.1 weeks and mean BW of 845 ± 265 grams, yielding 4387 SaO2–SpO2 datapoints. SpO2 overestimation, measured by mean bias, was 2.4-fold greater for Black infants and resulted in greater occult hypoxemia (SpO2 > 90% when SaO2 < 85%; 9.2% vs. 7.7% of samples). Sensitivity and specificity for detection of true hypoxemia were similar between groups (39 vs. 38%; 81 vs. 78%). Conclusion: There is a modest but consistent difference in SpO2 error between Black and White infants, with increased incidence of occult hypoxemia in Black infants.
AB - Objective: Pulse oximetry is commonly used in Neonatology, however recent adult data suggest racial disparity in accuracy, with overestimation of oxygen saturation for Black patients. Study design: Black and White infants <32 weeks gestation underwent simultaneous arterial blood gas and pulse oximetry measurement. Error by race was examined using mean bias, Arms, Bland–Altman, and linear/non-linear analysis. Results: A total of 294 infants (124 Black, 170 White) were identified with mean GA of 25.8 ± 2.1 weeks and mean BW of 845 ± 265 grams, yielding 4387 SaO2–SpO2 datapoints. SpO2 overestimation, measured by mean bias, was 2.4-fold greater for Black infants and resulted in greater occult hypoxemia (SpO2 > 90% when SaO2 < 85%; 9.2% vs. 7.7% of samples). Sensitivity and specificity for detection of true hypoxemia were similar between groups (39 vs. 38%; 81 vs. 78%). Conclusion: There is a modest but consistent difference in SpO2 error between Black and White infants, with increased incidence of occult hypoxemia in Black infants.
UR - http://www.scopus.com/inward/record.url?scp=85116946641&partnerID=8YFLogxK
U2 - 10.1038/s41372-021-01230-3
DO - 10.1038/s41372-021-01230-3
M3 - Article
C2 - 34642469
AN - SCOPUS:85116946641
SN - 0743-8346
VL - 42
SP - 79
EP - 85
JO - Journal of Perinatology
JF - Journal of Perinatology
IS - 1
ER -