Prospects for quantitative two-dimensional radiochromic film dosimetry for low dose-rate brachytherapy sources

Yi Le, Imad Ali, James F. Dempsey, Jeffrey F. Williamson

Research output: Contribution to journalArticlepeer-review

11 Scopus citations

Abstract

Radiochromic film (RCF) has been shown to be a precise and accurate two-dimensional dosimeter for acute exposure radiation fields. However, "temporal history" mismatch between calibration and brachytherapy films due to RCF dose-rate effects could introduce potentially large uncertainties in low dose-rate (LDR) brachytherapy absolute dose measurement. This article presents a quantitative evaluation of the precision and accuracy of a laser scanner-based RCF-dosimetry system and the effect of the temporal history mismatch in LDR absolute dose measurement. MD-55-2 RCF was used to measure absolute dose for a low dose-rate 137Cs brachytherapy source using both single- and double-exposure techniques. Dose-measurement accuracy was evaluated by comparing RCF to Monte Carlo photon-transport simulation. The temporal history mismatch effect was investigated by examining dependence of RCF accuracy on irradiation-to-densitometry time interval. The predictions of the empirical cumulative dose superposition model (CDSM) were compared with measurements. For the double-exposure technique, the agreement between measurement and Monte Carlo simulation was better than 4% in the 3-60 Gy dose range with measurement precisions (coverage factor k=1) of <2% and <6% for the doses greater or less than 3 Gy, respectively. The overall uncertainty (k=1) of dose rate/air-kerma strength measurements achievable by this dosimetry system for a spatial resolution of 0.1 mm is less than 4% for doses greater than 5 Gy. The measured temporal history mismatch systematic error is about 1.8% for a 48 h postexposure time when using the double exposure technique and agrees with CDSM's prediction qualitatively. This work demonstrates that the model MD-55-2 RCF detector has the potential to support quantitative dose measurements about LDR brachytherapy sources with precision and accuracy better than that of previously described dosimeters. The impacts of this work on the future use of new type of RCF were also discussed.

Original languageEnglish
Article number031612MPH
Pages (from-to)4622-4634
Number of pages13
JournalMedical physics
Volume33
Issue number12
DOIs
StatePublished - 2006

Keywords

  • Brachytherapy
  • Dosimetry
  • Low dose-rate
  • Monte Carlo
  • Radiochromic film (RCF)

Fingerprint

Dive into the research topics of 'Prospects for quantitative two-dimensional radiochromic film dosimetry for low dose-rate brachytherapy sources'. Together they form a unique fingerprint.

Cite this