T1 - Product ion distributions using H3O+ proton-transfer-reaction time-of-flight mass spectrometry (PTR-ToF-MS)
T2 - mechanisms, transmission effects, and instrument-to-instrument variability
AU - Link, Michael F.
AU - Claflin, Megan S.
AU - Cecelski, Christina E.
AU - Akande, Ayomide A.
AU - Kilgour, Delaney
AU - Heine, Paul A.
AU - Coggon, Matthew
AU - Stockwell, Chelsea E.
AU - Jensen, Andrew
AU - Yu, Jie
AU - Huynh, Han N.
AU - Ditto, Jenna C.
AU - Warneke, Carsten
AU - Dresser, William
AU - Gemmell, Keighan
AU - Jorga, Spiro
AU - Robertson, Rileigh L.
AU - de Gouw, Joost
AU - Bertram, Timothy
AU - Abbatt, Jonathan P.D.
AU - Borduas-Dedekind, Nadine
AU - Poppendieck, Dustin
N1 - Publisher Copyright:
© Author(s) 2025.
PY - 2025/2/27
Y1 - 2025/2/27
N2 - Proton-transfer-reaction mass spectrometry (PTR-MS) using hydronium ion (H3O+) ionization is widely used for the measurement of volatile organic compounds (VOCs) both indoors and outdoors. H3O+ ionization, as well as the associated chemistry in an ion–molecule reactor, is known to generate product ion distributions (PIDs) that include other product ions besides the proton-transfer product. We present a method, using gas-chromatography pre-separation, for quantifying PIDs from PTR-MS measurements of nearly 100 VOCs of different functional types including alcohols, ketones, aldehydes, acids, aromatics, organohalides, and alkenes. We characterize instrument configuration effects on PIDs and find that reactor reduced electric field strength (E/N), ion optic voltage gradients, and quadrupole settings have the strongest impact on measured PIDs. Through an interlaboratory comparison of PIDs measured from calibration cylinders, we characterized the variability of PID production from the same model of PTR-MS across seven participating laboratories. Product ion variability was generally smaller (e.g., < 20 %) for ions with larger contributions to the PIDs (e.g., > 0.30) but less predictable for product ions formed through O+2 and NO+ reactions. We present a publicly available library of H3O+ PTR-MS PIDs that will be updated periodically with user-provided data for the continued investigation into instrument-to-instrument variability of PIDs.
AB - Proton-transfer-reaction mass spectrometry (PTR-MS) using hydronium ion (H3O+) ionization is widely used for the measurement of volatile organic compounds (VOCs) both indoors and outdoors. H3O+ ionization, as well as the associated chemistry in an ion–molecule reactor, is known to generate product ion distributions (PIDs) that include other product ions besides the proton-transfer product. We present a method, using gas-chromatography pre-separation, for quantifying PIDs from PTR-MS measurements of nearly 100 VOCs of different functional types including alcohols, ketones, aldehydes, acids, aromatics, organohalides, and alkenes. We characterize instrument configuration effects on PIDs and find that reactor reduced electric field strength (E/N), ion optic voltage gradients, and quadrupole settings have the strongest impact on measured PIDs. Through an interlaboratory comparison of PIDs measured from calibration cylinders, we characterized the variability of PID production from the same model of PTR-MS across seven participating laboratories. Product ion variability was generally smaller (e.g., < 20 %) for ions with larger contributions to the PIDs (e.g., > 0.30) but less predictable for product ions formed through O+2 and NO+ reactions. We present a publicly available library of H3O+ PTR-MS PIDs that will be updated periodically with user-provided data for the continued investigation into instrument-to-instrument variability of PIDs.
UR - http://www.scopus.com/inward/record.url?scp=85219229845&partnerID=8YFLogxK
U2 - 10.5194/amt-18-1013-2025
DO - 10.5194/amt-18-1013-2025
M3 - Article
AN - SCOPUS:85219229845
SN - 1867-1381
VL - 18
SP - 1013
EP - 1038
JO - Atmospheric Measurement Techniques
JF - Atmospheric Measurement Techniques
IS - 4
ER -