T1 - Potential for localized, adjustable deep heating in soft-tissue environments with a 30-beam ultrasonic hyperthermia system
AU - Nussbaum, G. H.
AU - Straube, W. L.
AU - Drag, M. D.
AU - Melson, G. L.
AU - Emami, B.
AU - Sathiaseelan, V.
AU - Seppi, E.
AU - Shapiro, E.
PY - 1991/1/1
Y1 - 1991/1/1
N2 - Initial heating rates (°C/min) along parallel tracks at depths of 1-14 cm in a static, muscle-like phantom were determined from time-temperature profiles obtained with 'Helios', a 30-beam ultrasonic hyperthermia system developed by Varian Associates. Data were taken at a single operating frequency of 556 kHz, for different sets of focal plane ring diameters of the four-ring array applicator, different levels of transducer driving power and two different focal plane depths, 6 cm and 9 cm. In each experiment, at each point of temperature measurement, analysis of temperature versus time data over a 2 min heating interval permitted separation of the desired phantom heating from artefactual heating resulting primarily from absorption of transverse (shear) waves produced at phantom-metal probe catheter interfaces. The results of the studies conducted suggest that in a non-translating carriage mode, Helios can produce axially and laterally localized deep heating in soft tissues for tissue volumes of lateral dimension up to a minimum of 4 cm and tissue depths of at least 11 cm. The results obtained also suggest that Helios can produce laterally localized heating to tissue depths of at least 11 cm without excessive heating of superficial soft-tissue layers, for tissue volumes of lateral dimension up to a minimum of 8 cm. The methodology used in the phantom studies was applied to the production of localized heating in the right lobe of the liver of adult pigs. Temperature versus time profiles obtained in the in vivo studies indicated that, for the set of system parameters employed, concentration of ultrasonic power at greater depths in the liver (e.g. 10.5 cm versus 5 cm) could be achieved, suggesting that Helios should be able to produce localized heating of targeted hepatic volumes when its operating parameters are selected in accordance with effective treatment planning techniques.
AB - Initial heating rates (°C/min) along parallel tracks at depths of 1-14 cm in a static, muscle-like phantom were determined from time-temperature profiles obtained with 'Helios', a 30-beam ultrasonic hyperthermia system developed by Varian Associates. Data were taken at a single operating frequency of 556 kHz, for different sets of focal plane ring diameters of the four-ring array applicator, different levels of transducer driving power and two different focal plane depths, 6 cm and 9 cm. In each experiment, at each point of temperature measurement, analysis of temperature versus time data over a 2 min heating interval permitted separation of the desired phantom heating from artefactual heating resulting primarily from absorption of transverse (shear) waves produced at phantom-metal probe catheter interfaces. The results of the studies conducted suggest that in a non-translating carriage mode, Helios can produce axially and laterally localized deep heating in soft tissues for tissue volumes of lateral dimension up to a minimum of 4 cm and tissue depths of at least 11 cm. The results obtained also suggest that Helios can produce laterally localized heating to tissue depths of at least 11 cm without excessive heating of superficial soft-tissue layers, for tissue volumes of lateral dimension up to a minimum of 8 cm. The methodology used in the phantom studies was applied to the production of localized heating in the right lobe of the liver of adult pigs. Temperature versus time profiles obtained in the in vivo studies indicated that, for the set of system parameters employed, concentration of ultrasonic power at greater depths in the liver (e.g. 10.5 cm versus 5 cm) could be achieved, suggesting that Helios should be able to produce localized heating of targeted hepatic volumes when its operating parameters are selected in accordance with effective treatment planning techniques.
KW - Deep heating
KW - Heating patterns
KW - Hyperthermia
KW - Ultrasound
UR - http://www.scopus.com/inward/record.url?scp=0025804056&partnerID=8YFLogxK
U2 - 10.3109/02656739109004997
DO - 10.3109/02656739109004997
M3 - Article
C2 - 1880456
AN - SCOPUS:0025804056
SN - 0265-6736
VL - 7
SP - 279
EP - 299
JO - International Journal of Hyperthermia
JF - International Journal of Hyperthermia
IS - 2