Abstract

We present a frequency-domain near-infrared optical tomography system designed for breast cancer detection, in conjunction, with conventional ultrasound. It features fast optical switching, threewavelength excitations, and avalanche photodiode as detectors. Laser diodes at 660, 780, and 830 nm are used as light sources and their outputs are distributed sequentially to one of nine source fibers. An equivalent 130-dB isolation between electrical signals from different source channels is achieved with the optical switches of very low crosstalk. Ten detection channels, each of which includes a silicon avalanche photodiode, detect diffusive photon density waves simultaneously. The dynamic range of an avalanche photodiode is about 20 to 30 dB higher than that of a photomultiplier tube, thus eliminating the need for multistep system gain control. The entire system is compact in size (<0.051 m3) and fast in data acquisition (less than 2 sec for a complete scan). Calibration and the clinical experiment results are presented in the paper.

Original languageEnglish
Pages (from-to)504-510
Number of pages7
JournalJournal of biomedical optics
Volume9
Issue number3
DOIs
StatePublished - May 2004

Keywords

  • Avalanche photodiodes
  • Breast cancer
  • Diffusive imaging
  • Dynamic range
  • Optical switches

Fingerprint

Dive into the research topics of 'Portable near-infrared diffusive light imager for breast cancer detection'. Together they form a unique fingerprint.

Cite this