Abstract

An apparatus capable of measuring small fractional changes in ionization current has been used to study the effect of small inhomogeneities on photon dose in water. Small ring shaped inhomogeneities were introduced into a water phantom and measurements have been made for 4, 6, and 18 MV x rays. The results show (1) beyond the range of secondary electrons, the dose perturbation is basically a photon transport phenomenon which becomes less important as the beam energy increases; (2) within the range of secondary electrons, dose perturbation also involves electron transport, which has a strong dependence on atomic number and could result in a substantially large effect on dose deposition.

Original languageEnglish
Pages (from-to)78-83
Number of pages6
JournalMedical physics
Volume14
Issue number1
DOIs
StatePublished - Jan 1987

Keywords

  • PHANTOMS
  • RADIATION DOSE DISTRIBUTIONS
  • WATER
  • X-RAY DOSIMETRY