Performance evaluation of the microPET P4: A PET system dedicated to animal imaging

Y. C. Tai, A. Chatziioannou, S. Siegel, J. Young, D. Newport, R. N. Goble, R. E. Nutt, S. R. Cherry

Research output: Contribution to journalArticlepeer-review

282 Scopus citations

Abstract

The microPET Primate 4-ring system (P4) is an animal PET tomograph with a 7.8 cm axial extent, a 19 cm diameter transaxial field of view (FOV) and a 22 cm animal port. The system is composed of 168 detector modules, each with an 8 × 8 array of 2.2 × 2.2 × 10 mm3 lutetium oxyorthosilicate crystals, arranged as 32 crystal rings 26 cm in diameter. The detector crystals are coupled to a Hamamatsu R5900-C8 PS-PMT via a 10 cm long optical fibre bundle. The detectors have a timing resolution of 3.2 ns, an average energy resolution of 26%, and an average intrinsic spatial resolution of 1.75 mm. The system operates in 3D mode without inter-plane septa, acquiring data in list mode. The reconstructed image spatial resolution ranges from 1.8 mm at the centre to 3 mm at 4 cm radial offset. The tomograph has a peak system sensitivity of 2.25% at the centre of the FOV with a 250-750 keV energy window. The noise equivalent count rate peaks at 100-290 kcps for representative object sizes. Images from two phantoms and three different types of laboratory animal demonstrate the advantage of the P4 system over the original prototype microPET, including its threefold improvement in sensitivity and a large axial FOV sufficient to image an entire mouse in a single bed position.

Original languageEnglish
Pages (from-to)1845-1862
Number of pages18
JournalPhysics in medicine and biology
Volume46
Issue number7
DOIs
StatePublished - 2001

Fingerprint

Dive into the research topics of 'Performance evaluation of the microPET P4: A PET system dedicated to animal imaging'. Together they form a unique fingerprint.

Cite this