TY - JOUR
T1 - Performance assessment of a NaI(Tl) gamma counter for PET applications with methods for improved quantitative accuracy and greater standardization
AU - Lodge, Martin A.
AU - Holt, Daniel P.
AU - Kinahan, Paul E.
AU - Wong, Dean F.
AU - Wahl, Richard L.
N1 - Funding Information:
This work was supported by grants from the National Cancer Institute (U01CA140204, U01CA148131). PEK is one of the inventors of the Ge/Ga calibration system and may receive compensation from the patent licensing agreement between the University of Washington and RadQual, LLC. 68 68
Publisher Copyright:
© 2015, Lodge et al.; licensee Springer.
PY - 2015/12/1
Y1 - 2015/12/1
N2 - Background: Although NaI(Tl) gamma counters play an important role in many quantitative positron emission tomography (PET) protocols, their calibration for positron-emitting samples has not been standardized across imaging sites. In this study, we characterized the operational range of a gamma counter specifically for positron-emitting radionuclides, and we assessed the role of traceable 68Ge/68Ga sources for standardizing system calibration. Methods: A NaI(Tl) gamma counter was characterized with respect to count rate performance, adequacy of detector shielding, system stability, and sample volume effects using positron-emitting radionuclides (409- to 613-keV energy window). System efficiency was measured using 18F and compared with corresponding data obtained using a long-lived 68Ge/68Ga source that was implicitly traceable to a national standard. Results: One percent count loss was measured at 450 × 103 counts per minute. Penetration of the detector shielding by 511-keV photons gave rise to a negligible background count rate. System stability tests showed a coefficient of variation of 0.13% over 100 days. For a sample volume of 4 mL, the efficiencies relative to those at 0.1 mL were 0.96, 0.94, 0.91, 0.78, and 0.72 for 11C, 18F, 125I, 99mTc, and 51Cr, respectively. The efficiency of a traceable 68Ge/68Ga source was 30.1% ± 0.07% and was found to be in close agreement with the efficiency for 18F after consideration of the different positron fractions. Conclusions: Long-lived 68Ge/68Ga reference sources, implicitly traceable to a national metrology institute, can aid standardization of gamma counter calibration for 18F. A characteristic feature of positron emitters meant that accurate calibration could be maintained over a wide range of sample volumes by using a narrow energy window centered on the 511-keV peak.
AB - Background: Although NaI(Tl) gamma counters play an important role in many quantitative positron emission tomography (PET) protocols, their calibration for positron-emitting samples has not been standardized across imaging sites. In this study, we characterized the operational range of a gamma counter specifically for positron-emitting radionuclides, and we assessed the role of traceable 68Ge/68Ga sources for standardizing system calibration. Methods: A NaI(Tl) gamma counter was characterized with respect to count rate performance, adequacy of detector shielding, system stability, and sample volume effects using positron-emitting radionuclides (409- to 613-keV energy window). System efficiency was measured using 18F and compared with corresponding data obtained using a long-lived 68Ge/68Ga source that was implicitly traceable to a national standard. Results: One percent count loss was measured at 450 × 103 counts per minute. Penetration of the detector shielding by 511-keV photons gave rise to a negligible background count rate. System stability tests showed a coefficient of variation of 0.13% over 100 days. For a sample volume of 4 mL, the efficiencies relative to those at 0.1 mL were 0.96, 0.94, 0.91, 0.78, and 0.72 for 11C, 18F, 125I, 99mTc, and 51Cr, respectively. The efficiency of a traceable 68Ge/68Ga source was 30.1% ± 0.07% and was found to be in close agreement with the efficiency for 18F after consideration of the different positron fractions. Conclusions: Long-lived 68Ge/68Ga reference sources, implicitly traceable to a national metrology institute, can aid standardization of gamma counter calibration for 18F. A characteristic feature of positron emitters meant that accurate calibration could be maintained over a wide range of sample volumes by using a narrow energy window centered on the 511-keV peak.
KW - Calibration
KW - Efficiency
KW - Gamma counter
KW - Positron emission tomography
KW - Sample volume
KW - Standardization
KW - Well counter
UR - http://www.scopus.com/inward/record.url?scp=84957651322&partnerID=8YFLogxK
U2 - 10.1186/s40658-015-0114-3
DO - 10.1186/s40658-015-0114-3
M3 - Article
C2 - 26046011
AN - SCOPUS:84957651322
SN - 2197-7364
VL - 2
SP - 1
EP - 13
JO - EJNMMI Physics
JF - EJNMMI Physics
IS - 1
M1 - 11
ER -