TY - JOUR
T1 - Performance and acute procedural outcomes of the EnSite Precision™ cardiac mapping system for electrophysiology mapping and ablation procedures
T2 - results from the EnSite Precision™ observational study
AU - Hsu, Jonathan C.
AU - Darden, Douglas
AU - Glover, Benedict M.
AU - Colley, B. Judson
AU - Steinberg, Christian
AU - Thibault, Bernard
AU - Jewell, Coty
AU - Bernard, Michael
AU - Tabereaux, Paul B.
AU - Siddiqui, Usman
AU - Li, Jingyun
AU - Horvath, Eric E.
AU - Cooper, Daniel
AU - Lin, David
N1 - Publisher Copyright:
© 2022, The Author(s).
PY - 2022/10
Y1 - 2022/10
N2 - Background: The EnSite Precision™ cardiac mapping system (Abbott) is a catheter navigation and mapping system capable of displaying the three-dimensional (3D) position of conventional and sensor-enabled electrophysiology catheters, as well as displaying cardiac electrical activity as waveform traces and dynamic 3D maps of cardiac chambers. The EnSite Precision™ Observational Study (NCT-03260244) was designed to quantify and characterize the use of the EnSite Precision™ cardiac mapping system for mapping and ablation of cardiac arrhythmias in a real-world environment and evaluate procedural outcomes. Methods: A total of 1065 patients were enrolled at 38 centers in the USA and Canada between 2017 and 2018 and were followed for 12 months post procedure for arrhythmia recurrence, medication use, and quality-of-life changes. Eligible subjects were adults undergoing a cardiac electrophysiology mapping and radiofrequency ablation procedure using the EnSite Precision™ System. Results: A final cohort of 925 patients (64.3 years of age, 30.2% female) were analyzed. The primary procedural indication was atrial flutter in 48.1% (445/925), atrial fibrillation in 46.5% (430/925), and other arrhythmias in 5% (50/925). Electroanatomic mapping was performed in 81.5% (754/925) of patients. Mapping was stable throughout 79.8% (738/925) of procedures with initial mapping time of 8.6 min (IQR 4.7–15.0). Average mapping efficiency created with AutoMap or TurboMap was 164.9 ± 365.7 used points per minute. Median number of mapping points collected and used was 1752.5 and 811.0, respectively. Only 335/925 (36.2%) required editing and 66.0% (221/335) of these patients required editing of less than 10 points. Fluoroscopy was utilized in most cases (n = 811/925, 87.4%) with fluoroscopy time of 11.0 min (IQR 6.0–18.0). Overall median procedure time was 101.0 min (IQR 59.0–152.0). Acute procedural success was high for both atrial fibrillation (n = 422/430, 98.1%) and atrial flutter (n = 434/445, 97.5%). Conclusion: In a real-world study analysis, use of the EnSite Precision™ mapping system was associated with high procedural stability, short mapping times, high point density requiring infrequent editing, low fluoroscopy time, and high prevalence of acute procedural success.
AB - Background: The EnSite Precision™ cardiac mapping system (Abbott) is a catheter navigation and mapping system capable of displaying the three-dimensional (3D) position of conventional and sensor-enabled electrophysiology catheters, as well as displaying cardiac electrical activity as waveform traces and dynamic 3D maps of cardiac chambers. The EnSite Precision™ Observational Study (NCT-03260244) was designed to quantify and characterize the use of the EnSite Precision™ cardiac mapping system for mapping and ablation of cardiac arrhythmias in a real-world environment and evaluate procedural outcomes. Methods: A total of 1065 patients were enrolled at 38 centers in the USA and Canada between 2017 and 2018 and were followed for 12 months post procedure for arrhythmia recurrence, medication use, and quality-of-life changes. Eligible subjects were adults undergoing a cardiac electrophysiology mapping and radiofrequency ablation procedure using the EnSite Precision™ System. Results: A final cohort of 925 patients (64.3 years of age, 30.2% female) were analyzed. The primary procedural indication was atrial flutter in 48.1% (445/925), atrial fibrillation in 46.5% (430/925), and other arrhythmias in 5% (50/925). Electroanatomic mapping was performed in 81.5% (754/925) of patients. Mapping was stable throughout 79.8% (738/925) of procedures with initial mapping time of 8.6 min (IQR 4.7–15.0). Average mapping efficiency created with AutoMap or TurboMap was 164.9 ± 365.7 used points per minute. Median number of mapping points collected and used was 1752.5 and 811.0, respectively. Only 335/925 (36.2%) required editing and 66.0% (221/335) of these patients required editing of less than 10 points. Fluoroscopy was utilized in most cases (n = 811/925, 87.4%) with fluoroscopy time of 11.0 min (IQR 6.0–18.0). Overall median procedure time was 101.0 min (IQR 59.0–152.0). Acute procedural success was high for both atrial fibrillation (n = 422/430, 98.1%) and atrial flutter (n = 434/445, 97.5%). Conclusion: In a real-world study analysis, use of the EnSite Precision™ mapping system was associated with high procedural stability, short mapping times, high point density requiring infrequent editing, low fluoroscopy time, and high prevalence of acute procedural success.
KW - Catheter ablation
KW - Electroanatomical mapping
KW - Ensite mapping system
KW - Image processing
UR - http://www.scopus.com/inward/record.url?scp=85129766860&partnerID=8YFLogxK
U2 - 10.1007/s10840-022-01239-4
DO - 10.1007/s10840-022-01239-4
M3 - Article
C2 - 35536500
AN - SCOPUS:85129766860
SN - 1383-875X
VL - 65
SP - 141
EP - 151
JO - Journal of Interventional Cardiac Electrophysiology
JF - Journal of Interventional Cardiac Electrophysiology
IS - 1
ER -