Optode positional calibration in diffuse optical tomography

Jonathan J. Stott, Joseph P. Culver, Simon R. Arridge, David A. Boas

Research output: Contribution to journalArticlepeer-review

36 Scopus citations

Abstract

Although diffuse optical tomography is a highly promising technique used to noninvasively image blood volume and oxygenation, the reconstructed data are sensitive to systemic differences between the forward model and the actual experimental conditions. In particular, small changes in optode location or in the optode−tissue coupling coefficient significantly degrade the quality of the reconstruction images. Accurate system calibration therefore is an essential part of any experimental protocol. We present a technique for simultaneously calibrating optode positions and reconstructing images that significantly improves image quality, as we demonstrate with simulations and phantom experiments.

Original languageEnglish
Pages (from-to)3154-3162
Number of pages9
JournalApplied Optics
Volume42
Issue number16
DOIs
StatePublished - Jun 1 2003

Fingerprint

Dive into the research topics of 'Optode positional calibration in diffuse optical tomography'. Together they form a unique fingerprint.

Cite this