Abstract
We develope a method to optimize the resolution of diffuse optical tomographic instruments. Singular-value analysis of the tomographic weight matrix associated with specific data types, geometries, and optode arrangements is shown to provide a measure of image resolution. We achieve optimization of device configuration by monitoring the resolution measure described. We introduce this idea and demonstrate its utility by optimizing the spatial sampling interval and field-of-view parameters in the parallel-plane transmission geometry employed for diffuse optical breast imaging. We also compare resolution in transmission and remission geometries.
Original language | English |
---|---|
Pages (from-to) | 701-703 |
Number of pages | 3 |
Journal | Optics Letters |
Volume | 26 |
Issue number | 10 |
DOIs | |
State | Published - May 15 2001 |