Operator learning effect and interoperator reproducibility of the scanning laser polarimeter with variable corneal compensation

Shahar Frenkel, Efrat Slonim, Amjad Horani, Michal Molcho, Israel Barzel, Eytan Z. Blumenthal

Research output: Contribution to journalArticlepeer-review

26 Scopus citations

Abstract

Purpose: To ascertain operator learning effect, and to quantify the interoperator reproducibility, using the newly introduced GDx VCC (variable corneal compensator) scanning laser polarimeter. Design: Prospective instrument validation study. Participants: Three operators with no prior experience in operating the GDx VCC examined one randomly chosen eye of each of 30 randomly ordered subjects (15 glaucoma patients and 15 normal subjects). Methods: Each study eye was scanned by the 3 operators in a random sequence during a single session. Five GDx parameters were analyzed: TSNIT (temporal, superior, nasal, inferior, temporal) Average, TSNIT Standard Deviation (SD), Superior Average, Inferior Average, and Nerve Fiber Indicator (NFI). Main Outcome Measures: Retinal nerve fiber layer (RNFL) thickness GDx parameters across operators. Results: A learning effect was not found for any of the operators for any of the 5 GDx parameters studied. Reproducibility for each GDx parameter was measured as the mean SD value of the measurements taken by the 3 operators. Interoperator reproducibilities for normal and glaucomatous eyes, respectively, were 0.8 and 1.6 μm (TSNIT Average), 1.7 and 2.2 μm (Superior Average), 1.6 and 2.7 μm (Inferior Average), 1.3 and 1.7 μm (TSNIT SD), and 1.4 and 4.6 μm (NFI). The data obtained by each operator were highly correlated with those of the other 2 operators. Conclusions: With the commercially available GDx VCC, a learning effect was not found for 3 novice operators. In addition, RNFL measurements seem highly reproducible across operators.

Original languageEnglish
Pages (from-to)257-261
Number of pages5
JournalOphthalmology
Volume112
Issue number2
DOIs
StatePublished - Feb 1 2005

Fingerprint

Dive into the research topics of 'Operator learning effect and interoperator reproducibility of the scanning laser polarimeter with variable corneal compensation'. Together they form a unique fingerprint.

Cite this