MR safety and imaging of neuroform stents at 3T

  • Arvind Nehra
  • , Christopher J. Moran
  • , De Witte T. Cross
  • , Colin P. Derdeyn

Research output: Contribution to journalArticlepeer-review

28 Scopus citations

Abstract

The Neuroform stent is a self-expanding nitinol stent designed for use in wide-necked intracranial aneurysms. Heating and imaging artifacts were evaluated by using a porcine carotid artery aneurysm model in a 3T MR system. A suspended Neuroform stent was tested for deflection. No heating was measured, and no evidence of deflection of the stent was found. Imaging artifacts were minimal. MR imaging in patients with Neuroform stent-treated aneurysms is safe and feasible.

Original languageEnglish
Pages (from-to)1476-1478
Number of pages3
JournalAmerican Journal of Neuroradiology
Volume25
Issue number9
StatePublished - Oct 1 2004

Fingerprint

Dive into the research topics of 'MR safety and imaging of neuroform stents at 3T'. Together they form a unique fingerprint.

Cite this