T1 - Measurement and significance of antibodies against GM1ganglioside Report of a workshop, 18 April 1989, Chicago, IL, U.S.A.
AU - Marcus, Donald M.
AU - Latov, Norman
AU - Hsi, Bart P.
AU - Gillard, Baiba K.
AU - Antel, Jack
AU - Appel, Stanley
AU - Callegaro, Lanfranco
AU - Daune, Genevieve
AU - Latoy, Norman
AU - Marcus, Donald
AU - Nobile-Orazio, Eduardo
AU - Pestronk, Alan
AU - Portoukalian, Jacques
AU - Quarles, Richard
AU - Salazar-Grueso, Edgar
AU - Schneider, Larry
AU - Shy, Michael
AU - Steck, Andreas
PY - 1989/1/1
Y1 - 1989/1/1
N2 - Twelve laboratories from the United States, Canada, France, Italy and Switzerland participated in a workshop to compare assay used to measure anti-GM1antibodies, and to discuss the clinical significance of these antibodies. A panel of test samples containing varying amounts of anti-GM1antibody was prepared by mixing varied proportions of normal serum with a serum containing a monoclonal IgM antibody that bound GM1ganglioside. Enzyme-linked immunosorbent assay (ELISA) data were supplied by eight laboratories and ten laboratories classified the sera as negative, weakly of strongly positive. Most laboratories correctly identified the two samples that contained the highest quantities of antibody, but there was considerable disagreement on the classification of the three samples with moderate or small.
AB - Twelve laboratories from the United States, Canada, France, Italy and Switzerland participated in a workshop to compare assay used to measure anti-GM1antibodies, and to discuss the clinical significance of these antibodies. A panel of test samples containing varying amounts of anti-GM1antibody was prepared by mixing varied proportions of normal serum with a serum containing a monoclonal IgM antibody that bound GM1ganglioside. Enzyme-linked immunosorbent assay (ELISA) data were supplied by eight laboratories and ten laboratories classified the sera as negative, weakly of strongly positive. Most laboratories correctly identified the two samples that contained the highest quantities of antibody, but there was considerable disagreement on the classification of the three samples with moderate or small.
KW - Anti-ganglioside
KW - Enzyme-linked immunosorbent assay
KW - Neuropathy
UR - http://www.scopus.com/inward/record.url?scp=0024455146&partnerID=8YFLogxK
U2 - 10.1016/0165-5728(89)90144-6
DO - 10.1016/0165-5728(89)90144-6
M3 - Article
C2 - 2584398
AN - SCOPUS:0024455146
SN - 0165-5728
VL - 25
SP - 255
EP - 259
JO - Journal of Neuroimmunology
JF - Journal of Neuroimmunology
IS - 2-3
ER -