Low Rates of Aseptic Tibial Loosening in Obese Patients With Use of High-Viscosity Cement and Standard Tibial Tray: 2-Year Minimum Follow-Up

David A. Crawford, Keith R. Berend, Denis Nam, Robert L. Barrack, Joanne B. Adams, Adolph V. Lombardi

Research output: Contribution to journalArticlepeer-review

24 Scopus citations

Abstract

Background Total knee arthroplasty is overall a very successful surgery, but complications do occur. These complications include aseptic loosening of the tibial component, and obese patients are among the highest risk group. High-viscosity cement (HVC) has been implicated as a possible cause for aseptic loosening of the tibial component. The purpose of this study was to evaluate the incidence of aseptic loosening of the tibial component in obese patients with the use of HVC and standard tibial tray. Methods We identified 1366 obese patients (1851 knees) with a body mass index >35 kg/m2 and 2-year minimum follow-up who underwent primary total knee arthroplasty using HVC and a symmetrical, grit-blasted, cobalt-chrome tibial component with 40-mm stem. Preoperative and postoperative range of motion, Knee Society (KS) scores, complications, and reoperations were evaluated. Specifically, we assessed the rate of tibial aseptic loosening. Results At a mean 5.4 years follow-up, only 1 in 1851 knees had aseptic loosening of the tibial component for an incidence of 0.054%. There was a mean increase of 3.3 degrees of knee range of motion. KS pain level decreased by 38.6 points (50 point scale). KS clinical scores improved by 52.2, Knee Society functional scores improved by 19.5, University of California, Los Angeles, activity score improved by 0.9, and Oxford Knee Score by 15.7. All these improvements were statistically significant with P <.001. Conclusion Standard tibial components and HVC can be used in most patients, including the high-risk obese group, with low rates of tibial aseptic loosening.

Original languageEnglish
Pages (from-to)S183-S186
JournalJournal of Arthroplasty
Volume32
Issue number9
DOIs
StatePublished - Sep 2017

Keywords

  • aseptic
  • high viscosity
  • loosening
  • obese
  • tibia

Fingerprint

Dive into the research topics of 'Low Rates of Aseptic Tibial Loosening in Obese Patients With Use of High-Viscosity Cement and Standard Tibial Tray: 2-Year Minimum Follow-Up'. Together they form a unique fingerprint.

Cite this