TY - JOUR
T1 - Leadless pacemaker implantation under direct visualization during valve surgery
AU - Shivamurthy, Poojita
AU - Miller, Marc A.
AU - El-Eshmawi, Ahmed
AU - Boateng, Percy
AU - Pandis, Dimosthenis
AU - Pawale, Amit
AU - Leviner, Dror B.
AU - Costa, Ana Claudia
AU - Rimsukcharoenchai, Chartaroon
AU - Weiner, Menachem M.
AU - Salter, Benjamin
AU - Montgomery, Morgan L.
AU - Anyanwu, Anelechi
AU - Adams, David H.
N1 - Publisher Copyright:
© 2020 The American Association for Thoracic Surgery
PY - 2022/5
Y1 - 2022/5
N2 - Objective: The leadless cardiac pacemaker is typically implanted percutaneously and has been widely used for patients who have already undergone valve surgery. We sought to determine the feasibility and safety of implanting the leadless pacemaker under direct visualization during valve surgery. Methods: We performed a retrospective analysis of consecutive adult patients (n = 15) who underwent implantation of a leadless pacemaker under direct visualization at the time of valve surgery. Indications for single-chamber pacing were sick sinus syndrome with pauses (53.3%), atrial fibrillation with slow ventricular rates (13.3%) or complete heart block (6.6%), and elevated risk for postoperative heart block (26.6%). Leadless pacemaker performance and pacing percentage were assessed. Results: Patients' age was 67.5 ± 17 years, 6 patients (40%) were male, and 14 patients (93%) had atrial fibrillation. Isolated tricuspid valve replacement was performed in 5 patients (33.3%), and the remainder underwent multivalve surgery that included concomitant tricuspid valve repair/replacement. In 93% of the patients (n = 14), the immediate post–cardiopulmonary bypass pacing thresholds were normal (≤2.0 V at 0.24 ms) and normalized in the remaining patient by the next morning. The impedance/sensing values were normal and stable through follow-up (151 ± 119 days) in all patients. Reliable leadless pacemaker performance allowed for deferral of temporary epicardial wires in 11 patients (73%). There were no procedural complications or device malfunction. Conclusions: Leadless cardiac pacemaker implantation during valve surgery is feasible and safe. This hybrid approach to pacing may simplify the perioperative management of patients undergoing valve surgery who have an indication for single-chamber pacing.
AB - Objective: The leadless cardiac pacemaker is typically implanted percutaneously and has been widely used for patients who have already undergone valve surgery. We sought to determine the feasibility and safety of implanting the leadless pacemaker under direct visualization during valve surgery. Methods: We performed a retrospective analysis of consecutive adult patients (n = 15) who underwent implantation of a leadless pacemaker under direct visualization at the time of valve surgery. Indications for single-chamber pacing were sick sinus syndrome with pauses (53.3%), atrial fibrillation with slow ventricular rates (13.3%) or complete heart block (6.6%), and elevated risk for postoperative heart block (26.6%). Leadless pacemaker performance and pacing percentage were assessed. Results: Patients' age was 67.5 ± 17 years, 6 patients (40%) were male, and 14 patients (93%) had atrial fibrillation. Isolated tricuspid valve replacement was performed in 5 patients (33.3%), and the remainder underwent multivalve surgery that included concomitant tricuspid valve repair/replacement. In 93% of the patients (n = 14), the immediate post–cardiopulmonary bypass pacing thresholds were normal (≤2.0 V at 0.24 ms) and normalized in the remaining patient by the next morning. The impedance/sensing values were normal and stable through follow-up (151 ± 119 days) in all patients. Reliable leadless pacemaker performance allowed for deferral of temporary epicardial wires in 11 patients (73%). There were no procedural complications or device malfunction. Conclusions: Leadless cardiac pacemaker implantation during valve surgery is feasible and safe. This hybrid approach to pacing may simplify the perioperative management of patients undergoing valve surgery who have an indication for single-chamber pacing.
KW - intraoperative pacer
KW - leadless pacemaker
KW - pacemaker
KW - tricuspid valve
KW - valvular surgery
UR - http://www.scopus.com/inward/record.url?scp=85089885997&partnerID=8YFLogxK
U2 - 10.1016/j.jtcvs.2020.07.092
DO - 10.1016/j.jtcvs.2020.07.092
M3 - Article
C2 - 32891452
AN - SCOPUS:85089885997
SN - 0022-5223
VL - 163
SP - 1818
EP - 1825
JO - Journal of Thoracic and Cardiovascular Surgery
JF - Journal of Thoracic and Cardiovascular Surgery
IS - 5
ER -