Abstract
The dependencies of the light output response of CsI(Tl) crystals for various charged particle beams are investigated. Measurements were performed using 5.5MeV 241Am alpha particles, 220MeV alpha and 110MeV deuteron beams from the K500 Cyclotron at Texas A&M University, and p, 2H, 3He, 6Li, and 7Be beams from the Coupled Cyclotron Facility at Michigan State University. These measurements reveal a clear correlation between the non-uniformities observed in the same crystal for the various beams. Particle dependent global corrections were applied, resulting in final light output uniformities of the order of 0.1%. Annealed and non-annealed CsI(Tl) detectors and crystals with different thalium dopings are compared to determine whether annealing reduces the local variations in the light output. No significant correlation with the crystal fabrication process is observed.
| Original language | English |
|---|---|
| Pages (from-to) | 455-476 |
| Number of pages | 22 |
| Journal | Nuclear Instruments and Methods in Physics Research, Section A: Accelerators, Spectrometers, Detectors and Associated Equipment |
| Volume | 526 |
| Issue number | 3 |
| DOIs | |
| State | Published - Jul 1 2004 |
Keywords
- CsI(Tl) detectors
- Resolution
- Scintillator non-uniformity
- Scintillators - light output
Fingerprint
Dive into the research topics of 'Investigations and corrections of the light output uniformity of CsI(Tl) crystals'. Together they form a unique fingerprint.Cite this
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver