Abstract
Intratester test-retest reliability of insertion gain was determined on 28 subjects using the Frye 6500 real-ear analyzer. Results revealed mean differences of less than 1 dB for repeat measurements at six test frequencies between 250 and 4000 Hz. Also, over 80% of the repeat measures were within ±3 dB of the first measure. The results of this study are compared with previous studies of intratester reliability of insertion gain measures.
Original language | English |
---|---|
Pages (from-to) | 181-184 |
Number of pages | 4 |
Journal | Ear and hearing |
Volume | 11 |
Issue number | 3 |
DOIs | |
State | Published - Jun 1990 |