Integration of Real-Time Internal Electromagnetic Position Monitoring Coupled With Dynamic Multileaf Collimator Tracking: An Intensity-Modulated Radiation Therapy Feasibility Study

  • Ryan L. Smith
  • , Amit Sawant
  • , Lakshmi Santanam
  • , Raghu B. Venkat
  • , Laurence J. Newell
  • , Byung chul Cho
  • , Per Poulsen
  • , Herbert Catell
  • , Paul J. Keall
  • , Parag J. Parikh

Research output: Contribution to journalArticlepeer-review

39 Scopus citations

Abstract

Purpose: Continuous tumor position measurement coupled with a tumor tracking system would result in a highly accurate radiation therapy system. Previous internal position monitoring systems have been limited by fluoroscopic radiation dose and low delivery efficiency. We aimed to incorporate a continuous, electromagnetic, three-dimensional position tracking system (Calypso 4D Localization System) with a dynamic multileaf collimator (DMLC)-based dose delivery system. Methods and Materials: A research version of the Calypso System provided real-time position of three Beacon transponders. These real-time three-dimensional positions were sent to research MLC controller with a motion-tracking algorithm that changed the planned leaf sequence. Electromagnetic transponders were embedded in a solid water film phantom that moved with patient lung trajectories while being irradiated with two different plans: a step-and-shoot intensity-modulated radiation therapy (S-IMRT) field and a dynamic IMRT (D-IMRT) field. Dosimetric results were recorded under three conditions: no intervention, DMLC tracking, and a spatial gating system. Results: Dosimetric accuracy was comparable for gating and DMLC tracking. Failure rates for gating/DMLC tracking are as follows: ±3 cGy 10.9/ 7.5% for S-IMRT, 3.3/7.2% for D-IMRT; gamma (3mm/3%) 0.2/1.2% for S-IMRT, 0.2/0.2% for D-IMRT. DMLC tracking proved to be as efficient as standard delivery, with a two- to fivefold efficiency increase over gating. Conclusions: Real-time target position information was successfully integrated into a DMLC effector system to modify dose delivery. Experimental results show both comparable dosimetric accuracy as well as improved efficiency compared with spatial gating.

Original languageEnglish
Pages (from-to)868-875
Number of pages8
JournalInternational Journal of Radiation Oncology Biology Physics
Volume74
Issue number3
DOIs
StatePublished - Jul 1 2009