How image quality affects determination of target displacement when using kilovoltage cone-beam computed tomography

Seungjong Oh, Siyong Kim, Tae Suk Suh

Research output: Contribution to journalArticlepeer-review

11 Scopus citations

Abstract

The advent of kilovoltage cone-beam computed tomography (CBCT) integrated with a linear accelerator allows for more accurate image-guided radiotherapy (IGRT). The IGRT technique corrects target displacement based on internal body information obtained by acquiring the CBCT image set and registering it with the simulation CT image set just before the beam is delivered. In the present paper, we compare registration results by CBCT reconstruction quality (either high or medium). We analyzed data from a total of 56 CBCT projections from 6 patients. The translational vector differences were within 1 mm in all but 3 cases. The rotational displacement differences considered components of all three axes, and in 3 of 168 cases (56 projections by 3 axes), showed more than 1 degree of difference.

Original languageEnglish
Pages (from-to)101-107
Number of pages7
JournalJournal of applied clinical medical physics
Volume8
Issue number1
DOIs
StatePublished - 2007

Keywords

  • Cone-beam CT
  • IGRT
  • Reconstruction resolution
  • Target displacement

Fingerprint

Dive into the research topics of 'How image quality affects determination of target displacement when using kilovoltage cone-beam computed tomography'. Together they form a unique fingerprint.

Cite this