T1 - Final Two-Year Outcomes for the Sentry Bioconvertible Inferior Vena Cava Filter in Patients Requiring Temporary Protection from Pulmonary Embolism
AU - SENTRY Trial Investigators
AU - Dake, Michael D.
AU - Murphy, Timothy P.
AU - Krämer, Albrecht H.
AU - Darcy, Michael D.
AU - Sewall, Luke E.
AU - Curi, Michael A.
AU - Johnson, Matthew S.
AU - Arena, Frank
AU - Swischuk, James L.
AU - Ansel, Gary M.
AU - Silver, Mitchell J.
AU - Saddekni, Souheil
AU - Brower, Jayson S.
AU - Mendes, Robert
AU - Feezor, Robert
AU - Kalva, Sanjeeva
AU - Kies, Darren
AU - Bosiers, Marc
AU - Ziegler, Werner
AU - Farber, Mark
AU - Paolini, David
AU - Spillane, Robert
AU - Jones, Steven
AU - Peeters, Patrick
N1 - Publisher Copyright:
© 2019 SIR
PY - 2020/2
Y1 - 2020/2
N2 - Purpose: To report final 2-year outcomes with the Sentry bioconvertible inferior vena cava (IVC) filter in patients requiring temporary protection against pulmonary embolism (PE). Materials and Methods: In a prospective multicenter trial, the Sentry filter was implanted in 129 patients with documented deep vein thrombosis (DVT) and/or PE (67.5%) or who were at temporary risk of developing DVT/PE (32.6%). Patients were monitored and bioconversion status ascertained by radiography, computed tomography (CT), and CT venography through 2 years. Results: The composite primary 6-month endpoint of clinical success was achieved in 97.4% (111/114) of patients. The rate of new symptomatic PE was 0% (n = 126) through 1 year and 2.4% (n = 85) through the second year of follow-up, with 2 new nonfatal cases at 581 and 624 days that were adjudicated as not related to the procedure or device. Two patients (1.6%) developed symptomatic caval thrombosis during the first month and underwent successful interventions without recurrence. No other filter-related symptomatic complications occurred through 2 years. There was no filter tilting, migration, embolization, fracture, or caval perforation and no filter-related deaths through 2 years. Filter bioconversion was successful for 95.7% (110/115) of patients at 6 months, 96.4% (106/110) of patients at 12 months, and 96.5% (82/85) of patients at 24 months. Through 24 months of follow-up, there was no evidence of late-stage IVC obstruction or thrombosis after filter bioconversion or of thrombogenicity associated with retracted filter arms. Conclusions: The Sentry IVC filter provided safe and effective protection against PE, with a high rate of intended bioconversion and a low rate of device-related complications, through 2 years of follow-up.
AB - Purpose: To report final 2-year outcomes with the Sentry bioconvertible inferior vena cava (IVC) filter in patients requiring temporary protection against pulmonary embolism (PE). Materials and Methods: In a prospective multicenter trial, the Sentry filter was implanted in 129 patients with documented deep vein thrombosis (DVT) and/or PE (67.5%) or who were at temporary risk of developing DVT/PE (32.6%). Patients were monitored and bioconversion status ascertained by radiography, computed tomography (CT), and CT venography through 2 years. Results: The composite primary 6-month endpoint of clinical success was achieved in 97.4% (111/114) of patients. The rate of new symptomatic PE was 0% (n = 126) through 1 year and 2.4% (n = 85) through the second year of follow-up, with 2 new nonfatal cases at 581 and 624 days that were adjudicated as not related to the procedure or device. Two patients (1.6%) developed symptomatic caval thrombosis during the first month and underwent successful interventions without recurrence. No other filter-related symptomatic complications occurred through 2 years. There was no filter tilting, migration, embolization, fracture, or caval perforation and no filter-related deaths through 2 years. Filter bioconversion was successful for 95.7% (110/115) of patients at 6 months, 96.4% (106/110) of patients at 12 months, and 96.5% (82/85) of patients at 24 months. Through 24 months of follow-up, there was no evidence of late-stage IVC obstruction or thrombosis after filter bioconversion or of thrombogenicity associated with retracted filter arms. Conclusions: The Sentry IVC filter provided safe and effective protection against PE, with a high rate of intended bioconversion and a low rate of device-related complications, through 2 years of follow-up.
UR - http://www.scopus.com/inward/record.url?scp=85075473196&partnerID=8YFLogxK
U2 - 10.1016/j.jvir.2019.08.036
DO - 10.1016/j.jvir.2019.08.036
M3 - Article
C2 - 31711748
AN - SCOPUS:85075473196
SN - 1051-0443
VL - 31
SP - 221-230.e3
JO - Journal of Vascular and Interventional Radiology
JF - Journal of Vascular and Interventional Radiology
IS - 2
ER -