T1 - Experimental demonstration of benchtop x-ray fluorescence computed tomography (XFCT) of gold nanoparticle-loaded objects using lead- and tin-filtered polychromatic cone-beams
AU - Jones, Bernard L.
AU - Manohar, Nivedh
AU - Reynoso, Francisco
AU - Karellas, Andrew
AU - Cho, Sang Hyun
PY - 2012/12/7
Y1 - 2012/12/7
N2 - This report presents the first experimental demonstration, to our knowledge, of benchtop polychromatic cone-beam x-ray fluorescence computed tomography (XFCT) for a simultaneous determination of the spatial distribution and amount of gold nanoparticles (GNPs) within small-animal-sized objects. The current benchtop experimental setup successfully produced XFCT images accurately showing the regions containing small amount of GNPs (on the order of 0.1 mg) within a 3 cm diameter plastic phantom. In particular, the performance of the current XFCT setup was improved remarkably (e.g., at least a factor of 3 reduction in XFCT scan time) using a tin-filtered polychromatic beam in comparison with a lead-filtered beam. The results of this study strongly suggest that the current benchtop XFCT configuration can be made practical with a few modifications such as the deployment of array detectors, while meeting realistic constraints on x-ray dose, scan time and image resolution for routine pre-clinical in vivo imaging with GNPs.
AB - This report presents the first experimental demonstration, to our knowledge, of benchtop polychromatic cone-beam x-ray fluorescence computed tomography (XFCT) for a simultaneous determination of the spatial distribution and amount of gold nanoparticles (GNPs) within small-animal-sized objects. The current benchtop experimental setup successfully produced XFCT images accurately showing the regions containing small amount of GNPs (on the order of 0.1 mg) within a 3 cm diameter plastic phantom. In particular, the performance of the current XFCT setup was improved remarkably (e.g., at least a factor of 3 reduction in XFCT scan time) using a tin-filtered polychromatic beam in comparison with a lead-filtered beam. The results of this study strongly suggest that the current benchtop XFCT configuration can be made practical with a few modifications such as the deployment of array detectors, while meeting realistic constraints on x-ray dose, scan time and image resolution for routine pre-clinical in vivo imaging with GNPs.
UR - http://www.scopus.com/inward/record.url?scp=84870348240&partnerID=8YFLogxK
U2 - 10.1088/0031-9155/57/23/N457
DO - 10.1088/0031-9155/57/23/N457
M3 - Article
C2 - 23135315
AN - SCOPUS:84870348240
SN - 0031-9155
VL - 57
SP - N457-N467
JO - Physics in medicine and biology
JF - Physics in medicine and biology
IS - 23
ER -