Experience with the Amplatz retrievable vena caval filter. Work in progress

M. D. Darcy, J. F. Cardella, D. W. Hunter, T. P. Smith, W. R. Castaneda-Zuniga, G. Lund, K. Amplatz

Research output: Contribution to journalArticlepeer-review

20 Scopus citations

Abstract

The Amplatz retrievable vena caval filter was designed in an attempt to decrease complications associated with the placement of Mobin-Uddin or Kimray-Greenfield filters. The design allows percutaneous retrieval, thus expanding application of the filter to situations requiring temporary prophylaxis against pulmonary embolism. Filters have been placed in 16 patients, nine (56%) for prophylactic purposes. All filters were easily inserted percutaneously. Complications occurred in three patients; these included complete thrombosis of the inferior vena cava below the filter, misplacement of one filter into the pericaval retroperitoneal tissue, and development of thrombus cranial to the filter. With the current introduction system, the possibility of filter misplacement has been essentially eliminated. No patient experienced symptoms suggestive of pulmonary embolism after filter insertion. One filter retrieval has been performed, with no complications.

Original languageEnglish
Pages (from-to)611-614
Number of pages4
JournalRadiology
Volume161
Issue number3
DOIs
StatePublished - Jan 1 1986

Fingerprint

Dive into the research topics of 'Experience with the Amplatz retrievable vena caval filter. Work in progress'. Together they form a unique fingerprint.

Cite this