Abstract
Aims of the study: The safety and efficacy of a hemostatic powder (HP) versus a control agent, absorbable gelatin sponge and thrombin (G + T), were assessed, using a validated, quantitative bleeding severity scale. Methods: Subjects were randomized to receive HP (256 subjects) or G + T (132 subjects) for treatment of minimal, mild, or moderate bleeding at 20 investigational sites. The primary efficacy endpoint was non-inferiority of HP relative to G + T for success at achieving hemostasis within 6 minutes. Secondary endpoints in rank order included: superiority of HP relative to G + T in mean preparation time; non-inferiority of HP relative to G + T for achieving hemostasis within 3 min; superiority of HP relative to G + T for achieving hemostasis within 6 min; and superiority of HP relative to G + T for success for achieving hemostasis within 3 min. Results: A total of 388 subjects were included in the primary efficacy analysis. At 6 min, hemostasis was achieved in 93.0% (238/256) of the HP group compared to 77.3% (102/132) of the G + T group (non-inferiority P < 0.0001, superiority P < 0.0001). All secondary endpoints were met. Complications were comparable between treatment groups. Conclusions: HP had superior rates of hemostasis, shorter preparation time, and a similar safety profile compared to G + T in this prospective, randomized trial using quantitative bleeding severity criteria.
Original language | English |
---|---|
Pages (from-to) | 50-62 |
Number of pages | 13 |
Journal | Journal of cardiac surgery |
Volume | 34 |
Issue number | 1 |
DOIs | |
State | Published - Jan 2019 |
Keywords
- bleeding scale
- collagen
- hemostat
- hemostatic agent
- hemostatic powder
- thrombin