Evaluation of a new system for self blood glucose monitoring

  • William L. Clarke
  • , Dorothy J. Becker
  • , Daniel Cox
  • , Julio V. Santiago
  • , Neil H. White
  • , Jean Betschart
  • , Karen Eckenrode
  • , Lucy A. Levandoski
  • , Elizabeth A. Prusinski
  • , Linda M. Simineiro
  • , Andrea L. Snyder
  • , Ann M. Tideman
  • , Terri Yaeger

Research output: Contribution to journalArticlepeer-review

36 Scopus citations

Abstract

A new combination reflectance meter/visually interpretable system (Glucometer II®/Glucostix®, Ames Division, Miles Laboratories, Elkhart, IN) has been designed for self blood glucose monitoring. Performance evaluation of this system demonstrates a linear relationship between meter-determined blood glucose values and laboratory-determined whole blood glucose values (y = 0.95x + 2.86, r = 0.97). In addition, 95% of visually interpreted blood glucose values are within one color block of YSI comparative values. Error grid analysis, a new method for determining the clinical accuracy of patient-determined blood glucose results, demonstrated that components of this new system produce clinically accurate blood glucose results.

Original languageEnglish
Pages (from-to)209-213
Number of pages5
JournalDiabetes Research and Clinical Practice
Volume4
Issue number3
DOIs
StatePublished - Feb 19 1988

Keywords

  • Reflectance meter
  • Self blood glucose monitoring