Evaluation of a new system for self blood glucose monitoring

William L. Clarke, Dorothy J. Becker, Daniel Cox, Julio V. Santiago, Neil H. White, Jean Betschart, Karen Eckenrode, Lucy A. Levandoski, Elizabeth A. Prusinski, Linda M. Simineiro, Andrea L. Snyder, Ann M. Tideman, Terri Yaeger

Research output: Contribution to journalArticlepeer-review

36 Scopus citations

Abstract

A new combination reflectance meter/visually interpretable system (Glucometer II®/Glucostix®, Ames Division, Miles Laboratories, Elkhart, IN) has been designed for self blood glucose monitoring. Performance evaluation of this system demonstrates a linear relationship between meter-determined blood glucose values and laboratory-determined whole blood glucose values (y = 0.95x + 2.86, r = 0.97). In addition, 95% of visually interpreted blood glucose values are within one color block of YSI comparative values. Error grid analysis, a new method for determining the clinical accuracy of patient-determined blood glucose results, demonstrated that components of this new system produce clinically accurate blood glucose results.