TY - JOUR
T1 - Establishment of reference intervals for thyroid-stimulating hormone and free thyroxine in amniotic fluid using the Bayer ADVIA Centaur
AU - Baumann, Nikola A.
AU - Gronowski, Ann M.
PY - 2007/7
Y1 - 2007/7
N2 - Thyroid hormone concentrations in amniotic fluid (AF) can aid in the diagnosis of fetal hypothyroidism. However, the availability of reference intervals for thyroid hormones in AF using current methods is limited. The purpose of this study was to validate the Bayer ADVIA Centaur (Bayer Healthcare, Tarrytown, NY) third-generation thyroid-stimulating hormone (TSH), total thyroxine (T 4), and free thyroxine (FT 4) assays for use with AF and establish reference intervals for these thyroid hormones in AF. Third-trimester AF samples were assayed for TSH, T 4, and FT 4. Reference intervals were calculated using nonparametric analyses. The reference intervals were as follows: TSH (n = 131), 0.04 to 0.51 μIU/mL (0.04-0.51 mIU/L), with a median of 0.10 μIU/mL (0.10 mIU/L); FT 4 (n = 133), less than 0.10 to 0.77 ng/dL (1.29-9.93 pmol/L), with a median of 0.26 ng/dL (3.35 pmol/L). T 4 in AF was undetectable using the Centaur assay, reinforcing the importance of validating different manufacturers' immunoassays for use with nonserum specimens. These reference intervals represent the first study performed to date using a third-generation TSH assay and a sensitive FT 4 assay.
AB - Thyroid hormone concentrations in amniotic fluid (AF) can aid in the diagnosis of fetal hypothyroidism. However, the availability of reference intervals for thyroid hormones in AF using current methods is limited. The purpose of this study was to validate the Bayer ADVIA Centaur (Bayer Healthcare, Tarrytown, NY) third-generation thyroid-stimulating hormone (TSH), total thyroxine (T 4), and free thyroxine (FT 4) assays for use with AF and establish reference intervals for these thyroid hormones in AF. Third-trimester AF samples were assayed for TSH, T 4, and FT 4. Reference intervals were calculated using nonparametric analyses. The reference intervals were as follows: TSH (n = 131), 0.04 to 0.51 μIU/mL (0.04-0.51 mIU/L), with a median of 0.10 μIU/mL (0.10 mIU/L); FT 4 (n = 133), less than 0.10 to 0.77 ng/dL (1.29-9.93 pmol/L), with a median of 0.26 ng/dL (3.35 pmol/L). T 4 in AF was undetectable using the Centaur assay, reinforcing the importance of validating different manufacturers' immunoassays for use with nonserum specimens. These reference intervals represent the first study performed to date using a third-generation TSH assay and a sensitive FT 4 assay.
KW - Amniotic fluid
KW - Reference intervals
KW - Thyroid hormone
UR - http://www.scopus.com/inward/record.url?scp=34547092935&partnerID=8YFLogxK
U2 - 10.1309/69A5AV266W230AUA
DO - 10.1309/69A5AV266W230AUA
M3 - Article
C2 - 17580284
AN - SCOPUS:34547092935
SN - 0002-9173
VL - 128
SP - 158
EP - 163
JO - American journal of clinical pathology
JF - American journal of clinical pathology
IS - 1
ER -