Abstract
5 TABLE An example of generalized 2D/3D FSE VFL scan parameters for GYN and prostate BT from two institutions (Table presented.) 1Ax = Axial, Sag = Sagittal, and PA = Para-Axial, PS = Para-Sagittal, PC = Para-Coronal to Applicator for GYN 2Full 3D gradient non-linearity (GNL) correction may not be supported for 2D sequences. 3Effective TE reported for 3D FSE VFL 4Echo train duration reported for 3D FSE VFL 5Readout bandwidth reported for 1.5T/3.0T; Additional optimization to recover SNR may be required. 6Longer scan times may benefit from administration of antispasmodic agents to reduce motion. 7Use of in-plane and through plane interpolation and acceleration methods (e.g., partial Fourier and parallel imaging) can introduce blurring and artifacts and should be verified prior to clinical use.
Original language | English |
---|---|
Pages (from-to) | 1258 |
Number of pages | 1 |
Journal | Medical physics |
Volume | 50 |
Issue number | 2 |
DOIs |
|
State | Published - Feb 2023 |