TY - JOUR
T1 - Dosimetric impact of online correction via cone-beam ct-based image guidance for stereotactic lung radiotherapy
AU - Galerani, Ana Paula
AU - Grills, Inga
AU - Hugo, Geoffrey
AU - Kestin, Larry
AU - Mohammed, Nasiruddin
AU - Chao, K. Kenneth
AU - Suen, Andrew
AU - Martinez, Alvaro
AU - Yan, Di
PY - 2010/12/1
Y1 - 2010/12/1
N2 - Purpose: To evaluate the dosimetric impact of online cone-beam computed tomography (CBCT) guided correction in lung stereotactic body radiation therapy (SBRT). Methods and Materials: Twenty planning and 162 CBCT images from 20 patients undergoing lung SBRT were analyzed. The precorrection CBCT (CBCT after patient setup, no couch correction) was registered to planning CT using soft tissue; couch shift was applied, with a second CBCT for verification (postcorrection CBCT). Targets and normal structures were delineated on CBCTs: gross tumor volume (GTV), clinical target volume (CTV), cord, esophagus, lung, proximal bronchial tree, and aorta. Dose distributions on all organs manifested on each CBCT were compared with those planned on the CT. Results: Without CBCT guided target position correction, target dose reduced with respect to treatment plan. Mean and standard deviation of treatment dose discrepancy from the plan were -3.2% (4.9%), -2.1% (4.4%), -6.1% (10.7%), and -3.5% (7%) for GTV D 99%, GTV D95%, CTV D99%, and CTV D 95%, respectively. With CBCT correction, the results were -0.4% (2.6%), 0.1% (1.7%), -0.3% (4.2%), and 0.5% (3%). Mean and standard deviation of the difference in normal organ maximum dose were 2.2% (6.5%) before correction and 2.4% (5.9%) after correction for esophagus; 6.1% (14.1%) and 3.8% (8.1%) for cord; 3.1% (17.5%) and 6.2% (9.8%) for proximal bronchial tree; and 17.7% (19.5%) and 14.1% (17%) for aorta. Conclusion: Online CBCT guidance improves the accuracy of target dose delivery for lung SBRT. However, treatment dose to normal tissue can vary regardless of the correction. Normal tissues should be considered during target registration, according to target proximity.
AB - Purpose: To evaluate the dosimetric impact of online cone-beam computed tomography (CBCT) guided correction in lung stereotactic body radiation therapy (SBRT). Methods and Materials: Twenty planning and 162 CBCT images from 20 patients undergoing lung SBRT were analyzed. The precorrection CBCT (CBCT after patient setup, no couch correction) was registered to planning CT using soft tissue; couch shift was applied, with a second CBCT for verification (postcorrection CBCT). Targets and normal structures were delineated on CBCTs: gross tumor volume (GTV), clinical target volume (CTV), cord, esophagus, lung, proximal bronchial tree, and aorta. Dose distributions on all organs manifested on each CBCT were compared with those planned on the CT. Results: Without CBCT guided target position correction, target dose reduced with respect to treatment plan. Mean and standard deviation of treatment dose discrepancy from the plan were -3.2% (4.9%), -2.1% (4.4%), -6.1% (10.7%), and -3.5% (7%) for GTV D 99%, GTV D95%, CTV D99%, and CTV D 95%, respectively. With CBCT correction, the results were -0.4% (2.6%), 0.1% (1.7%), -0.3% (4.2%), and 0.5% (3%). Mean and standard deviation of the difference in normal organ maximum dose were 2.2% (6.5%) before correction and 2.4% (5.9%) after correction for esophagus; 6.1% (14.1%) and 3.8% (8.1%) for cord; 3.1% (17.5%) and 6.2% (9.8%) for proximal bronchial tree; and 17.7% (19.5%) and 14.1% (17%) for aorta. Conclusion: Online CBCT guidance improves the accuracy of target dose delivery for lung SBRT. However, treatment dose to normal tissue can vary regardless of the correction. Normal tissues should be considered during target registration, according to target proximity.
KW - Cone-beam computed tomography
KW - Dosimetry
KW - Image-guided radiotherapy
KW - Lung cancer
KW - Stereotactic radiotherapy
UR - http://www.scopus.com/inward/record.url?scp=78549281030&partnerID=8YFLogxK
U2 - 10.1016/j.ijrobp.2010.02.012
DO - 10.1016/j.ijrobp.2010.02.012
M3 - Article
C2 - 20646857
AN - SCOPUS:78549281030
SN - 0360-3016
VL - 78
SP - 1571
EP - 1578
JO - International Journal of Radiation Oncology Biology Physics
JF - International Journal of Radiation Oncology Biology Physics
IS - 5
ER -