Cylindrical lens configuration for optimizing light delivery in a curvilinear endocavity photoacoustic imaging system

  • Yixiao Lin
  • , Sitai Kou
  • , Yun Zou
  • , Konstantin Maslov
  • , Quing Zhu

Research output: Contribution to journalArticlepeer-review

2 Scopus citations

Abstract

Curvilinear endocavity ultrasound images capture a wide field of view with a miniature probe. In adapting photoacoustic imaging (PAI) to work with such ultrasound systems, light delivery is challenged by the trade-off between image quality and laser safety concerns. Here, we present two novel, to the best of our knowledge, designs based on cylindrical lenses that are optimized for transvaginal PAI B-scan imaging. Our simulation and experimental results demonstrate that, compared to conventional light delivery methods for PAI imaging, the proposed designs are safer for higher pulse energies and provide deeper imaging and a wider lateral field of view. The proposed designs could also improve the performance of endoscopic co-registered ultrasound/photoacoustic imaging in other clinical applications.

Original languageEnglish
Pages (from-to)2417-2420
Number of pages4
JournalOptics Letters
Volume48
Issue number9
DOIs
StatePublished - May 1 2023

Fingerprint

Dive into the research topics of 'Cylindrical lens configuration for optimizing light delivery in a curvilinear endocavity photoacoustic imaging system'. Together they form a unique fingerprint.

Cite this