Compensating for intraoperative soft-tissue deformations using incomplete surface data and finite elements

David M. Cash, Michael I. Miga, Tuhin K. Sinha, Robert L. Galloway, William C. Chapman

Research output: Contribution to journalArticlepeer-review

99 Scopus citations

Abstract

Image-guided liver surgery requires the ability to identify and compensate for soft tissue deformation in the organ. The predeformed state is represented as a complete three-dimensional surface of the organ, while the intraoperative data is a range scan point cloud acquired from the exposed liver surface. The first step is to rigidly align the coordinate systems of the intraoperative and preoperative data. Most traditional rigid registration methods minimize an error metric over the entire data set. In this paper, a new deformation-identifying rigid registration (DIRR) is reported that identifies and aligns minimally deformed regions of the data using a modified closest point distance cost function. Once a rigid alignment has been established, deformation is accounted for using a linearly elastic finite element model (FEM) and implemented using an incremental framework to resolve geometric nonlinearities. Boundary conditions for the incremental formulation are generated from intraoperatively acquired range scan surfaces of the exposed liver surface. A series of phantom experiments is presented to assess the fidelity of the DIRR and the combined DIRR/FEM approaches separately. The DIRR approach identified deforming regions in 90% of cases under conditions of realistic surgical exposure. With respect to the DIRR/FEM algorithm, subsurface target errors were correctly located to within 4 mm in phantom experiments.

Original languageEnglish
Pages (from-to)1479-1491
Number of pages13
JournalIEEE Transactions on Medical Imaging
Volume24
Issue number11
DOIs
StatePublished - Nov 2005

Keywords

  • Deformation identification
  • Finite element modeling
  • Organ deformation
  • Registration

Fingerprint

Dive into the research topics of 'Compensating for intraoperative soft-tissue deformations using incomplete surface data and finite elements'. Together they form a unique fingerprint.

Cite this