T1 - Comparison of Two Commercially Available Fentanyl Screening Immunoassays for Clinical Use
AU - Budelier, Melissa M.
AU - Franks, Caroline E.
AU - Logsdon, Nicole
AU - Jannetto, Paul J.
AU - Scott, Mitchell G.
AU - Roper, Stephen M.
AU - Farnsworth, Christopher W.
N1 - Publisher Copyright:
© American Association for Clinical Chemistry 2020. All rights reserved. For permissions, please email: [email protected].
Copyright:
This record is sourced from MEDLINE/PubMed, a database of the U.S. National Library of Medicine
PY - 2020/11
Y1 - 2020/11
N2 - Fentanyl is a synthetic opioid associated with illicit drug use and overdose deaths. The SEFRIA Immunalysis (IAL) and ARK fentanyl assays are both FDA-cleared, open channel immunoassays for fentanyl detection in urine. However, limited data are available in the literature comparing these assays. The objective of this study was to perform a direct comparison of these two fentanyl immunoassays. Methods: IAL and ARK fentanyl immunoassays were performed on a Roche Cobas e602 automated chemistry analyzer. Repeatability and total imprecision were compared by diluting fentanyl into urine at concentrations above, below, and at the manufacturers' cutoffs of 1.0 ng/mL. Cross-reactivity was assessed for norfentanyl and the fentanyl analogs acetylfentanyl, acrylfentanyl, and furanylfentanyl. Concordance was assessed in 90 patient samples using liquid chromatography-tandem mass spectrometry (LC-MS/MS) as the gold standard. Results: Repeatability varied from 11.4%-17.8% on the IAL assay and 2.8%-5.5% on the ARK assay. Total imprecision was 18.9%-40.7% on the IAL assay and 2.9%-6.4% on the ARK assay. Both assays cross-reacted with acetylfentanyl (∼100%), acrylfentanyl (∼100%), and furanylfentanyl (∼20%), but only the ARK assay cross-reacted with norfentanyl (∼3%). An admixture of 0.5 ng/mL fentanyl and 6 ng/mL norfentanyl produced a positive result on the ARK assay. Total concordance between IAL and ARK for 90 tested patient samples was 93% (kappa = 0.85). Relative to LC-MS/MS, the IAL assay had a concordance of 90% (kappa = 0.79) and the ARK assay had a concordance of 94% (kappa = 0.88). Including norfentanyl in the LC-MS/MS confirmation increased the concordance of the ARK to 96% (kappa = 0.90). Conclusions: The ARK assay recognized the metabolite norfentanyl, demonstrated superior precision, and had better concordance with LC-MS/MS compared to the IAL assay.
AB - Fentanyl is a synthetic opioid associated with illicit drug use and overdose deaths. The SEFRIA Immunalysis (IAL) and ARK fentanyl assays are both FDA-cleared, open channel immunoassays for fentanyl detection in urine. However, limited data are available in the literature comparing these assays. The objective of this study was to perform a direct comparison of these two fentanyl immunoassays. Methods: IAL and ARK fentanyl immunoassays were performed on a Roche Cobas e602 automated chemistry analyzer. Repeatability and total imprecision were compared by diluting fentanyl into urine at concentrations above, below, and at the manufacturers' cutoffs of 1.0 ng/mL. Cross-reactivity was assessed for norfentanyl and the fentanyl analogs acetylfentanyl, acrylfentanyl, and furanylfentanyl. Concordance was assessed in 90 patient samples using liquid chromatography-tandem mass spectrometry (LC-MS/MS) as the gold standard. Results: Repeatability varied from 11.4%-17.8% on the IAL assay and 2.8%-5.5% on the ARK assay. Total imprecision was 18.9%-40.7% on the IAL assay and 2.9%-6.4% on the ARK assay. Both assays cross-reacted with acetylfentanyl (∼100%), acrylfentanyl (∼100%), and furanylfentanyl (∼20%), but only the ARK assay cross-reacted with norfentanyl (∼3%). An admixture of 0.5 ng/mL fentanyl and 6 ng/mL norfentanyl produced a positive result on the ARK assay. Total concordance between IAL and ARK for 90 tested patient samples was 93% (kappa = 0.85). Relative to LC-MS/MS, the IAL assay had a concordance of 90% (kappa = 0.79) and the ARK assay had a concordance of 94% (kappa = 0.88). Including norfentanyl in the LC-MS/MS confirmation increased the concordance of the ARK to 96% (kappa = 0.90). Conclusions: The ARK assay recognized the metabolite norfentanyl, demonstrated superior precision, and had better concordance with LC-MS/MS compared to the IAL assay.
KW - Fentanyl
KW - drug screen
KW - method comparison
UR - http://www.scopus.com/inward/record.url?scp=85095802967&partnerID=8YFLogxK
U2 - 10.1093/jalm/jfaa048
DO - 10.1093/jalm/jfaa048
M3 - Article
C2 - 32674121
AN - SCOPUS:85095802967
SN - 2576-9456
VL - 5
SP - 1277
EP - 1286
JO - The journal of applied laboratory medicine
JF - The journal of applied laboratory medicine
IS - 6
ER -