T1 - Comparison of forehead and digit oximetry in surgical/trauma patients at risk for decreased peripheral perfusion
AU - Schallom, Lynn
AU - Sona, Carrie
AU - McSweeney, Maryellen
AU - Mazuski, John
N1 - Funding Information:
Study funded by Nellcor/Tyco Healthcare, Pleasanton, California.
PY - 2007/5
Y1 - 2007/5
N2 - Objective: Measurement of pulse oximetry (Spo2) is often impaired in critically ill patients. Forehead reflectance oximetry, the Max-Fast (Nellcor, Pleasanton, CA), may be less susceptible to poor tissue perfusion and could improve accuracy of oxygen saturation measurement. The objective of this study was to evaluate the use of forehead oximetry measures in critically ill surgical/trauma patients. Methods: A prospective interventional study of 30 critically ill surgical/trauma patients at risk for decreased peripheral perfusion, as evidenced by need for vasopressor support (24 patients), transfusion of more than 6 unit packed cells in 24 hours (two patients), or an inability to obtain consistent saturation from a digit sensor (four patients), compared forehead and digit-based oximeter Spo2 readings with co-oximetry (Sao2) measurements from arterial blood samples. Sao2 values were converted to functional oxygen saturation (SO2) measurements for the final comparison. Patients were fitted with forehead (Nellcor Max-Fast) and digit (Nellcor Max A; digit 1) sensors connected to Nellcor OxiMax N-595 oximeters and a digit sensor (Nellcor Max A; digit 2) connected to a multiparameter monitor (Philips CMS [Andover, MA]). Three measurements of Sao2 were obtained from each subject over a 24-hour time period, and simultaneous measurements of Spo2 were recorded from the three monitors. Results: The three Spo2 measurements (forehead, digit 1, and digit 2) were compared with SO2 values using the Bland-Altman method to assess agreement. Forehead measurements demonstrated a mean bias of -1.39, whereas digit 1 was -2.61 and digit 2 was -3.84. Pearson correlations (r) for forehead, digit 1, and digit 2 with SO2 were .834, .433, and .254, respectively. There were fewer unsuccessful measurements with the forehead oximetry technique. Conclusions: Forehead sensors improve measurement of oxygen saturation in critically ill surgical/trauma patients at risk for decreased peripheral perfusion.
AB - Objective: Measurement of pulse oximetry (Spo2) is often impaired in critically ill patients. Forehead reflectance oximetry, the Max-Fast (Nellcor, Pleasanton, CA), may be less susceptible to poor tissue perfusion and could improve accuracy of oxygen saturation measurement. The objective of this study was to evaluate the use of forehead oximetry measures in critically ill surgical/trauma patients. Methods: A prospective interventional study of 30 critically ill surgical/trauma patients at risk for decreased peripheral perfusion, as evidenced by need for vasopressor support (24 patients), transfusion of more than 6 unit packed cells in 24 hours (two patients), or an inability to obtain consistent saturation from a digit sensor (four patients), compared forehead and digit-based oximeter Spo2 readings with co-oximetry (Sao2) measurements from arterial blood samples. Sao2 values were converted to functional oxygen saturation (SO2) measurements for the final comparison. Patients were fitted with forehead (Nellcor Max-Fast) and digit (Nellcor Max A; digit 1) sensors connected to Nellcor OxiMax N-595 oximeters and a digit sensor (Nellcor Max A; digit 2) connected to a multiparameter monitor (Philips CMS [Andover, MA]). Three measurements of Sao2 were obtained from each subject over a 24-hour time period, and simultaneous measurements of Spo2 were recorded from the three monitors. Results: The three Spo2 measurements (forehead, digit 1, and digit 2) were compared with SO2 values using the Bland-Altman method to assess agreement. Forehead measurements demonstrated a mean bias of -1.39, whereas digit 1 was -2.61 and digit 2 was -3.84. Pearson correlations (r) for forehead, digit 1, and digit 2 with SO2 were .834, .433, and .254, respectively. There were fewer unsuccessful measurements with the forehead oximetry technique. Conclusions: Forehead sensors improve measurement of oxygen saturation in critically ill surgical/trauma patients at risk for decreased peripheral perfusion.
UR - http://www.scopus.com/inward/record.url?scp=34248150312&partnerID=8YFLogxK
U2 - 10.1016/j.hrtlng.2006.07.007
DO - 10.1016/j.hrtlng.2006.07.007
M3 - Article
C2 - 17509425
AN - SCOPUS:34248150312
SN - 0147-9563
VL - 36
SP - 188
EP - 194
JO - Heart and Lung: Journal of Acute and Critical Care
JF - Heart and Lung: Journal of Acute and Critical Care
IS - 3
ER -