TY - JOUR
T1 - Comparison between continuous non-invasive estimated cardiac output by pulse wave transit time and thermodilution method
AU - Sinha, Ashish C.
AU - Singh, Preet Mohinder
AU - Grewal, Navneet
AU - Aman, Mansoor
AU - Dubowitz, Gerald
PY - 2014/10/1
Y1 - 2014/10/1
N2 - Aims and Objectives: Cardiac output (CO) measurement is essential for many therapeutic decisions in anesthesia and critical care. Most available non-invasive CO measuring methods have an invasive component. We investigate "pulse wave transit time" (estimated continuous cardiac output [esCCO]) a method of CO measurement that has no invasive component to its use. Materials and Methods: After institutional ethical committee approval, 14 adult (21-85 years) patients undergoing surgery and requiring pulmonary artery catheter (PAC) for measuring CO, were included. Postoperatively CO readings were taken simultaneously with thermodilution (TD) via PAC and esCCO, whenever a change in CO was expected due to therapeutic interventions. Both monitoring methods were continued until patients' discharge from the Intensive Care Unit and observer recording values using TD method was blinded to values measured by esCCO system. Results: Three hundred and one readings were obtained simultaneously from both methods. Correlation and concordance between the two methods was derived using Bland-Altman analysis. Measured values showed significant correlation between esCCO and TD ( r = 0.6, P < 0.001, 95% confidence limits of 0.51-0.68). Mean and (standard deviation) for bias and precision were 0.13 (2.27) L/min and 6.56 (2.19) L/min, respectively. The 95% confidence interval for bias was - 4.32 to 4.58 L/min and for precision 2.27 to10.85 L/min. Conclusions: Although, esCCO is the only true non-invasive continuous CO monitor available and even though its values change proportionately to TD method (gold standard) with the present degree of error its utility for clinical/therapeutic decision-making is questionable.
AB - Aims and Objectives: Cardiac output (CO) measurement is essential for many therapeutic decisions in anesthesia and critical care. Most available non-invasive CO measuring methods have an invasive component. We investigate "pulse wave transit time" (estimated continuous cardiac output [esCCO]) a method of CO measurement that has no invasive component to its use. Materials and Methods: After institutional ethical committee approval, 14 adult (21-85 years) patients undergoing surgery and requiring pulmonary artery catheter (PAC) for measuring CO, were included. Postoperatively CO readings were taken simultaneously with thermodilution (TD) via PAC and esCCO, whenever a change in CO was expected due to therapeutic interventions. Both monitoring methods were continued until patients' discharge from the Intensive Care Unit and observer recording values using TD method was blinded to values measured by esCCO system. Results: Three hundred and one readings were obtained simultaneously from both methods. Correlation and concordance between the two methods was derived using Bland-Altman analysis. Measured values showed significant correlation between esCCO and TD ( r = 0.6, P < 0.001, 95% confidence limits of 0.51-0.68). Mean and (standard deviation) for bias and precision were 0.13 (2.27) L/min and 6.56 (2.19) L/min, respectively. The 95% confidence interval for bias was - 4.32 to 4.58 L/min and for precision 2.27 to10.85 L/min. Conclusions: Although, esCCO is the only true non-invasive continuous CO monitor available and even though its values change proportionately to TD method (gold standard) with the present degree of error its utility for clinical/therapeutic decision-making is questionable.
KW - Continuous cardiac output monitor
KW - Non-invasive cardiac output measurement
KW - Pulse wave transit time
UR - http://www.scopus.com/inward/record.url?scp=84907987561&partnerID=8YFLogxK
U2 - 10.4103/0971-9784.142059
DO - 10.4103/0971-9784.142059
M3 - Article
C2 - 25281622
AN - SCOPUS:84907987561
SN - 0971-9784
VL - 17
SP - 273
EP - 277
JO - Annals of Cardiac Anaesthesia
JF - Annals of Cardiac Anaesthesia
IS - 4
ER -