Comparison between continuous non-invasive estimated cardiac output by pulse wave transit time and thermodilution method

Ashish C. Sinha, Preet Mohinder Singh, Navneet Grewal, Mansoor Aman, Gerald Dubowitz

Research output: Contribution to journalArticlepeer-review

13 Scopus citations

Abstract

Aims and Objectives: Cardiac output (CO) measurement is essential for many therapeutic decisions in anesthesia and critical care. Most available non-invasive CO measuring methods have an invasive component. We investigate "pulse wave transit time" (estimated continuous cardiac output [esCCO]) a method of CO measurement that has no invasive component to its use. Materials and Methods: After institutional ethical committee approval, 14 adult (21-85 years) patients undergoing surgery and requiring pulmonary artery catheter (PAC) for measuring CO, were included. Postoperatively CO readings were taken simultaneously with thermodilution (TD) via PAC and esCCO, whenever a change in CO was expected due to therapeutic interventions. Both monitoring methods were continued until patients' discharge from the Intensive Care Unit and observer recording values using TD method was blinded to values measured by esCCO system. Results: Three hundred and one readings were obtained simultaneously from both methods. Correlation and concordance between the two methods was derived using Bland-Altman analysis. Measured values showed significant correlation between esCCO and TD ( r = 0.6, P < 0.001, 95% confidence limits of 0.51-0.68). Mean and (standard deviation) for bias and precision were 0.13 (2.27) L/min and 6.56 (2.19) L/min, respectively. The 95% confidence interval for bias was - 4.32 to 4.58 L/min and for precision 2.27 to10.85 L/min. Conclusions: Although, esCCO is the only true non-invasive continuous CO monitor available and even though its values change proportionately to TD method (gold standard) with the present degree of error its utility for clinical/therapeutic decision-making is questionable.

Original languageEnglish
Pages (from-to)273-277
Number of pages5
JournalAnnals of Cardiac Anaesthesia
Volume17
Issue number4
DOIs
StatePublished - Oct 1 2014

Keywords

  • Continuous cardiac output monitor
  • Non-invasive cardiac output measurement
  • Pulse wave transit time

Fingerprint

Dive into the research topics of 'Comparison between continuous non-invasive estimated cardiac output by pulse wave transit time and thermodilution method'. Together they form a unique fingerprint.

Cite this