TY - JOUR
T1 - Characterizing MR Imaging isocenter variation in MRgRT
AU - Kim, Taeho
AU - Gu, Bruce
AU - Maraghechi, Borna
AU - Green, Olga
AU - Lewis, Benjamin
AU - Mutic, Sasa
AU - Gach, H. Michael
N1 - Publisher Copyright:
© 2020 IOP Publishing Ltd.
PY - 2020
Y1 - 2020
N2 - We characterized MRI isocenter variation at various gantry positions in two 0.35 T MRgRT systems using two independent methods. First, image center-based quantification was employed on 3D volumetric and 2D cine images of a 24 cm diameter spherical phantom at various gantry positions in the MRI QA mode. The center of the phantom images was identified to quantify the variation of the imaging center at each gantry position. Second, image registration-based quantification was used in radiotherapy mode. 3D volumetric MRIs of a cylindrical phantom were acquired and corresponding image registration from MRI to planning CT was performed. The shifts of the couch were identified to quantify the variation of the imaging center. For verification of noticeable MRI isocenter variation, star-shot pattern measurements with five beams were delivered on the radio-chromic film inserted into the phantom after the couch was shifted. The center of the star-shot pattern was identified to quantify the variation of the imaging center. The proposed methods for measuring MRI isocenter variation were demonstrated with MR-LINAC and MR-60Co systems. Both of the MRgRT systems had field inhomogeneities <5 ppm over a 24 cm diameter spherical volume (DSV) and spatial integrity distortion: <1 mm within 100 mm radius and <2 mm within 175 mm radius. The MRI isocenter of the MR-LINAC system showed noticeable 3D variation (max magnitude: 1.8 mm) compared to that of MR-60Co system (max magnitude: 0.9 mm) relative to the reference gantry positions. In addition, 2D variations (max magnitude) of the MRI isocenter from sagittal cine images were 0.9 mm for the MR-LINAC system and 0.5 mm for the MR-60Co system. Two proposed methods quantified the MRI isocenter variation for various gantry positions in two 0.35 T MRgRT systems. The results of significant isocenter variation in the MR-LINAC system requires further investigation to determine the cause.
AB - We characterized MRI isocenter variation at various gantry positions in two 0.35 T MRgRT systems using two independent methods. First, image center-based quantification was employed on 3D volumetric and 2D cine images of a 24 cm diameter spherical phantom at various gantry positions in the MRI QA mode. The center of the phantom images was identified to quantify the variation of the imaging center at each gantry position. Second, image registration-based quantification was used in radiotherapy mode. 3D volumetric MRIs of a cylindrical phantom were acquired and corresponding image registration from MRI to planning CT was performed. The shifts of the couch were identified to quantify the variation of the imaging center. For verification of noticeable MRI isocenter variation, star-shot pattern measurements with five beams were delivered on the radio-chromic film inserted into the phantom after the couch was shifted. The center of the star-shot pattern was identified to quantify the variation of the imaging center. The proposed methods for measuring MRI isocenter variation were demonstrated with MR-LINAC and MR-60Co systems. Both of the MRgRT systems had field inhomogeneities <5 ppm over a 24 cm diameter spherical volume (DSV) and spatial integrity distortion: <1 mm within 100 mm radius and <2 mm within 175 mm radius. The MRI isocenter of the MR-LINAC system showed noticeable 3D variation (max magnitude: 1.8 mm) compared to that of MR-60Co system (max magnitude: 0.9 mm) relative to the reference gantry positions. In addition, 2D variations (max magnitude) of the MRI isocenter from sagittal cine images were 0.9 mm for the MR-LINAC system and 0.5 mm for the MR-60Co system. Two proposed methods quantified the MRI isocenter variation for various gantry positions in two 0.35 T MRgRT systems. The results of significant isocenter variation in the MR-LINAC system requires further investigation to determine the cause.
UR - http://www.scopus.com/inward/record.url?scp=85083091255&partnerID=8YFLogxK
U2 - 10.1088/2057-1976/ab7bc6
DO - 10.1088/2057-1976/ab7bc6
M3 - Article
C2 - 33438654
AN - SCOPUS:85083091255
SN - 2057-1976
VL - 6
JO - Biomedical Physics and Engineering Express
JF - Biomedical Physics and Engineering Express
IS - 3
M1 - 035009
ER -